COM Server Commands
32
Table 3.
Tr
: Trace Number
Data Type
Long
Description
Trace number
Range
from 1 to 8
Out of Range
An error occurs. Error code: 202.
Notes
If the trace number is not specified, it is taken as equal to 1.
SCPI.CALCulate(
Ch
).PARameter(
Tr
).SELect
Object Type
Method
Target
Trace
Tr
of channel
Ch
,
Ch
Tr
trace number 1–8 (see
Description
Sets the active channel and trace.
Syntax
app.SCPI.CALCulate(Ch).PARameter(Tr).SELect
Notes
If the channel number is greater than the number of the channels
displayed, an error occurs and the command is ignored. If the trace
number is greater than the number of the traces displayed in the
channel, an error occurs and the command is ignored.
Equivalent
Softkeys
Channel > Active Channel
Trace > Active Trace