INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS
6.11.5 Calibrate Value Zone-4/Zone-L (tn-4/tn-L)
This is only activated in the L126 Dual Workstation. The tn-4/tn-L option is for the calibration of
temperature on Zone-4/Zone-L, only one option is active (See Section 7: Different Heated Surfaces).
If there is an offset between the value on the display and any measurements made with a high
precision external temperature sensor, this can be corrected. The new calibration value will be kept
as the displayed value, with temperature control conducted on this basis (See Section 9: Temperature
Calibration).
Follow these steps to navigate in the Calibrate Zone-4/Zone-L (tn-4/tn-L) option.
ACTION
KEY
DISPLAY
Press and hold both the UP
and
DOWN key for 3 seconds to
enter the menu.
Press the DOWN key until the menu "StuP" appears in the
display.
Press the SP key once to enter the Setup menu.
Press the DOWN key until the option "tn-4" / "tn-L" appears in
the display.
or
Press and hold the SP key to see the current value.
While holding the SP key, press either the UP
or
DOWN key
to select the calibration value.
When the required value has been selected, release the SP
key.
Press and hold both the UP
and
DOWN keys for 3 seconds
to exit the menu.
NOTE:
Always wait for the temperature to be stabilized before calibrating.
Summary of Contents for K-Systems L100 IVF Series
Page 1: ...K SYSTEMS IVF L100 Series Workstations Instruction Manual...
Page 76: ...76 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 77: ...77 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 78: ...78 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 79: ...79 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...
Page 80: ...80 INSTRUCTION MANUAL FOR L100 SERIES WORKSTATIONS...