29
9
Settings
1.
CAUTION:
Changing the calibration value should only be done based on measurement
performed by an experienced user or a service technician authorized by CooperSurgical
2.
CAUTION:
Changing the regulation factor should only be done by a service technician
authorized by CooperSurgical
3.
CAUTION:
Do not use the incubator without the UV light. This setting should only be used by
service technicians
4.
CAUTION:
Remove all samples from the unit before decontamination, as the unit will be
heated to more than 60°C and the gas will be turned off
5.
CAUTION:
The incubator should not be used without gas; if the gas concentration is wrong the
embryos might be harmed or even perish. Gas may be turned off to reduce gas consumption if
the are no samples inside
NOTES
1.
NOTE:
The CO
2
regulator must be activated before concentration can be viewed or changed
2.
NOTE:
Gas concentration should be calibrated using an external precision gas analyzer
through the gas sampling outlet. Only use a gas analyzer recommended by CooperSurgical
3.
NOTE:
Always connect a 100% pure CO
2
source to the CO
2
gas inlet located on the back of
G185 incubator before switching the regulator on
4.
NOTE:
The O
2
regulator must be activated before the concentration can be viewed or changed
5.
NOTE:
Always connect a 100% N
2
source to the gas inlet located on the back of the incubator
before switching the regulator on
6.
NOTE:
CO
2
gas consumption is not a fixed value and may vary from 0 l/h – 10 l/h. The
necessary amount is automatically calculated and adjusted by the system
7.
NOTE:
100% pure and dry CO
2
gas will be needed for controllers to operate correctly
8.
NOTE:
N
2
gas consumption is not a fixed value and may vary from 0 l/h – 52 l/h. The necessary
amount is automatically calculated and adjusted by the system
9.
NOTE:
100% pure and dry N
2
will be needed to reduce the O
2
level in the system
10.
NOTE:
Having the parameter set incorrectly when no PC is attached will not make any
difference. Switching to “OFF” or “PDA” when the PC is attached corrupts the data
11.
NOTE:
For data logger installation, please refer to section 10 - “Section 10 - Data Logger” on
12.
NOTE:
Never attempt to switch off dataflow when the PC is logging data. Stop the program
on the PC first
13.
NOTE:
Calibration of temperature is done by adjusting the tn-x (where x is the zone number)
according to a measurement performed in both chambers within the zone by a precision
thermometer. Only use thermometers recommended by CooperSurgical
14.
NOTE:
After temperature adjustment, allow at least 15 minutes for the temperature to
stabilize. Use the thermometer to verify correct temperatures in each zone
15.
NOTE:
If one warmed area is much warmer than another in its proximity, there will be cross-
over heating. Calibration is performed best by starting with tn-1 and continuing to tn-2 to tn-6.
Then let the system stabilize and repeat the procedure if necessary