Cooper DTW 200 User Manual Download Page 2

CF 139 

DTW 200/200N (12/10) 

 

General Operation 

1.  Press and hold MODE for one second to select from the following measuring modes: 
 

Real Time – displays torque transients (no output or indicator) 
PP Mode – Peak, captures peak torque (peak data output, PP appears on the display) 
PD Mode – First Peak, capture first peak value (peak data output, PD appears on the display) 
C Mode – Continuous RS-232 output, Display and output torque transients (12 or 180 data/sec., C appears 
on the display. 

2.  Place the appropriate size socket onto the drive, and if desired, the one-way clutch also. 
3.  Rotate the wrench head so the display may be viewed while measuring. 
4.  Press the CLEAR button on the display unit to zero the display. Insert the socket onto the bolthead, hold 

perpendicularly (not at an angle) and turn to measure.  

5.  After measuring, press CLEAR to zero display for the next test. 

 

Programming 

Click ON/OFF to turn on. Press PRG for one second. The display shows HI and then 
the High setpoint value. This confirms the tester is ready for the following 
programming steps. 
 

1.  High Setpoint (HI) 

After HI is displayed, and the high setpoint value, press DOWN  to move the 
numeric place and press UP to select values, then press PRG to enter. 

2.  Low Setpoint (LO) 

After high value is entered, LO is displayed, then the low setpoint value. Press 
DOWN to move the numeric place and press UP to select values, then press 
PRG to enter. 

3.  Peak Down Minimum (PdLO) 

After low value is entered, PdLO is displayed, then the PdLO value. Press down 
to move the numeric value and press up to select values, and press PRG to enter.  

PdLO sets a minimum torque value for Peak Down mode. For example, if PdLO value is set at 5.0 lbf-in, only a 
reading over that will be measured in Peak Down mode. 
4.  Continuous Data Output Minimum 

After Peak Down value is entered, CLO is displayed, then the CLO value. . Press DOWN to move the numeric 
place and press UP to select values, then press PRG to enter. 

5.  Auto Zero Reset (AC) 

After Continuous Data Output Minimum is entered, AC is displayed and then the Auto Zero Reset duration 
value. Press Up or Down to select 0.0C-0.1C-0.5C-1.0C-1.5C-2.0C-2.5C-3.0C, and press PRG to enter (0.1C 
for 0.1 second and 0.0C for Manual Reset) 

6.  Batch Counter (CO) 

After the Zero Value is entered, CO is displayed then the CO value. Use the UP or DOWN keys to select a 
value and PRG to enter (0-99, clockwise only) 

7. Interval 

(In) 

After the Batch Counter value is entered, In is displayed then the In value. the UP or DOWN keys to select a 
value and PRG to enter (0-24 in 22-second increments) 