Installation
515 HC04 - 8 June 2007
17
Pulse Output Connection
Figure 11 shows a connection example for a pulse output. Output channel 1
uses terminals 27 (+) and 28 (-). Output channel 2 uses terminals 29 (+) and
30 (-).
Figure 11 Output Pulse Connection Diagram
Control Relays (Alarms)
The standard instrument has two alarm relays, which can be used to drive
external devices such as external relays, lamps, and audible alarms. The
advanced option has four alarm relays.
The operation of each alarm relay can be set to various modes as described
in
There is also an equipment failure alarm option. This alarm can have
normally closed (open) contacts which open (close) when the instrument
displays any error message as listed in
there is a loss of power to the instrument.
The output characteristics of the relays are:
Note:
Solid state relays use AC voltage only.
+
-
External Load
Resistor 10K
Vo +8...24V DC
15
Logic Input
Open
DC Ground
Collector
Output
16
Pulse Output
(OUTn)
Shield
18
500
Maximum Voltage
30 volts DC or 250 volts AC
Maximum Current
3A