CMS50DL1   

User Manual

 

P u l s e   O x i m e t e r  

 

 
Contec Medical Systems Co., Ltd. 

Address:No.112 Qinhuang West Street, Economic 

Technical 

Development Zone, Qinhuangdao, Hebei Province, PEOPLE

’S 

REPUBLIC OF CHINA 
Tel: +86-335-8015430 
Fax: +86-335-8015588 
Technical s86-335-8015431 
E-mail: [email protected] 
Website: http://www.contecmed.com

 

EC REPRESENTATIVE 

Shanghai International Holding Corp. GmbH(Europe) 

Address: Eiffestrasse 80, 20537, Hamburg, Germany   
Tel: 

49-40-2513175   

Fax: 

49-40-255726 

E-mail: [email protected]

 

CMS2.782.179(CE.USA)ESS/1.9    1.4.01.01.701    2021.02 

Instructions to User 

Dear  users,  thank  you  very  much  for  purchasing  the  Pulse  Oximeter  (hereinafter 

referred to as device). 

This  Manual  is  written  and  compiled  in  accordance  with  the

 

council  directive 

MDD93/42/EEC  for  medical  devices  and  harmonized  standards.  In  case  of 

modifications and  software  upgrades, the  information  contained  in this document is 

subject to change without notice. 

It is a medical device, which can be used repeatedly. 
The  Manual  describes,  in  accordance  with  the  device’s  features  and  requirements, 
main  structure,  functions,  specifications,  correct  methods  for  transportation, 

installation,  usage,  operation,  repair,  maintenance  and  storage,  etc.  as  well  as  the 

safety procedures to protect both the user and device. Refer to the respective chapters 

for details. 

Please  read  the  User  Manual  carefully  before  using  this  device.  The  User  Manual 

which  describes  the  operating  procedures  should  be  followed  strictly.  Failure  to 

follow  the  User  Manual  may  cause  measuring  abnormality,  device  damage  and 

human  injury.  The  manufacturer  is  NOT  responsible  for  the  safety,  reliability  and 

performance  issues  and  any  monitoring  abnormality,  human  injury  and  device 

damage  due  to  users'  negligence  of  the  operation  instructions.  The  manufacturer’s 

warranty service does not cover such faults. 

Owing to the forthcoming renovation, the specific products you received may not be 

totally  in accordance  with the  description  of  this  User  Manual.  We  would  sincerely 

regret for that. 

Our company has the final interpretation to this manual. The content of this manual is 

subject to change without prior notice. 

Warnings 

Remind that it may cause serious consequences to tester, user or environment. 

 

Explosive  hazard—DO  NOT  use  the  device  in  environment  with  inflammable 

gas such as anesthetic. 

 

DO NOT use the device while examining by MRI or CT, as the induced current 

may cause burn. 

 

Do not take the information displayed on the device as the sole basis for clinical 

diagnosis.  The  device  is  only  used  as  an  auxiliary  means  in  diagnosis.  And  it 
must  be  used  in  conjunction  with  doctor’s  advice,  clinical  manifestations  and 
symptoms. 

 

The maintenance to the device. Users are not permitted to maintain or refit the 

device by themselves. 

 

Uncomfortable  or  painful  feeling  may  appear  if  using  the  device  ceaselessly, 

especially for the microcirculation disturbance users. It is not recommended that 

the sensor is used on the same finger for more than 2 hours. 

 

For  some  special  users  who  need  a  more  careful  inspection  on  the  test  site, 
please don’t place the device on the edema or tender tissue. 

 

Please  do  not  stare  at  the  red  and  infrared  light  emitter  (the  infrared  light  is 

invisible) after turning on the device, including the maintenance staff, as it may 

be harmful to the eyes. 

 

The  device  contains  silicone,  PVC,  TPU,  TPE  and  ABS  materials,  whose 

biocompatibility  has  been  tested  in  accordance  with  the  requirements  in  ISO 

10993-1,  and  it has  passed  the recommended  biocompatibility  test.  The  person 

who is allergic to silicone, PVC, TPU, TPE or ABS can not use this device. 

 

Do  NOT  strand  the  lanyard  to  avoid  device  drop  and  damage.  The  lanyard  is 

made  of  insensitive  material.  Please  do  not  use  it  if  any  person  is  allergic  to 

lanyard. Do not wrap the lanyard around neck to avoid an accident. 

 

The  disposal  of  scrap  device,  its  accessories  and  packaging  should  follow  the 

local laws and regulations, to avoid polluting to the local environment. And the 

packaging materials must be placed in the region where the children are out of 

reaching. 

 

The  device  can not  be  used  with  the  equipment not  specified  in  the  Manual.  Only  the 

accessories  appointed  or  recommended  by  the  manufacturer  can  be  used,  otherwise  it 

may cause injury to the tester and operator or damage to the device. 

 

Check the device before use to make sure that there is no visible damage that may affect 
user’s safety and device performance. When there is obvious damage, please replace the 
damaged parts before use. 

 

Functional testers can not be used to assess the accuracy of the Pulse Oximeter. 

 

Some  functional testers  or  patient  simulators  can  be  used to  verify  whether the  device 

works normally, for example, INDEX-2LFE Simulator (software version: 3.00), please 

refer to the Manual for the detailed operation steps. 

 

Some  functional  testers  or  patient  simulators  can  measure  the  accuracy  of  the  device 

copied calibration curve, but they can not be used to evaluate the device accuracy. 

 

When  using  the  device,  please  keep  it  away  from  the  equipment  which  can  generate 

strong  electric  field  or  strong  magnetic  field.  Using  the  device  in  an  inappropriate 

environment  may  cause  interference  to  the  surrounding  radio  equipment  or  affect  its 

working. 

 

The measured accuracy will be affected by the interference of electrosurgical equipment. 

 

When several products are used on the same patient simultaneously, danger may occur 

which is arisen from the overlap of leakage current. 

 

CO  poisoning  will  appear  excessive  estimation,  so  it  is  not  recommended  to  use  the 

device. 

 

This device is not intended for treatment. 

 

The intended operator of the device may be a patient. 

 

Avoid maintaining the device during using. 

1

 Overview

 

The oxygen saturation is the percentage of HbO

2

 in the total Hb in the blood, so-called the O

2

 

concentration in the blood, it is an important physiological parameter for the respiratory and 

circulatory  system.  A  number  of  diseases  related  to  respiratory  system  may  cause  the 

decrease  of  SpO

2

  in  the  blood,  furthermore,  some  other  causes  such  as  the  malfunction  of 

human  body's  self-adjustment,  damages  during  surgery,  and  the  injuries  caused  by  some 

medical checkup would also lead to the difficulty of oxygen supply in human body, and the 

corresponding symptoms would appear as a consequence, such as vertigo, impotence,  vomit 

etc. Serious symptoms might bring danger to human's life. Therefore, prompt information of 

patients' SpO

2

 is of great help for the doctor to discover the potential danger, and is of great 

importance in the clinical medical field.   

Insert the finger when measuring, the device will directly display the SpO

2

 value measured, it 

has a higher accuracy and repeatability. 

1.1 Features   

A. Easy to use. 

B. Small in volume, light in weight, convenient to carry. 

C. Low power consumption. 

1.2 Applied range 

The  Pulse  Oximeter  can  be  used  to  measure  human  Hemoglobin  Saturation  and  pulse  rate 

through finger, and indicate the pulse intensity by the bar-display. The product is suitable for 

use  in  family,  hospital(Ordinary  sickroom  ),  Oxygen  Bar,  social  medical  organizations  and 

also the measure of saturation oxygen and pulse rate. 

1.3 Environment requirements 

Storage Environment 

a) Temperature: -40 

 ~ + 60 

 

b) Relative humidity: ≤ 95% 

c) Atmospheric pressure: 500 hPa ~ 1060 hPa 

Operating Environment 

a) Temperature: +10 

~ + 40 

 

b) Relative Humidity: ≤ 75% 

c) Atmospheric pressure: 700 hPa ~ 1060 hPa 

1.4 Precautions 

1.4.1 Attention 

Point out conditions or practices that may cause damage to the device or other properties. 

 

Before using the device, make sure that it locates in normal working state and operating 

environment. 

 

In  order  to  get  a  more  accurate  measurement,  it  should  be  used  in  a  quiet  and 

comfortable environment. 

 

When it is carried from cold environment to warm or humid environment, please do not 

use it immediately. 

 

If the device is splashed or coagulated by water, please stop operating. 

 

DO NOT operate the device with sharp things. 

 

High temperature, high pressure, gas sterilizing or immersion disinfection for the device 

is  not  permitted.  Refer  to  User  Manual  in  the  relative  chapter  (6.1)  for  cleaning  and 

disinfection..Please take out the internal battery before cleaning and disinfection. 

 

The device is suitable for children and adult. 

 

The device may not be suitable for all users, if you can't get a satisfactory result, please 

stop using it. 

 

Data averaging and signal processing have a delay in the upgrade of SpO

2

 data values. 

When  the  data  update  period  is  less  than  30  seconds,  the  time  for  obtaining  dynamic 

average values will increase, which is arisen from signal degradation, low perfusion or 

other interference, it depends on the PR value. 

 

The device has 3-year service life, date of manufacture: see the label. 

 

The  device  hasn't  low-voltage  prompt  function,  it  only  shows  the  low-voltage,  please 

change the battery when the battery voltage is used up. 

 

The maximum temperature at the SpO

2

 probe -tissue interface should be less than 41

 

which is measured by the temperature tester. 

 

During measuring, when abnormal conditions appear on the screen, please pull out your 

finger and reinsert it to measure again. 

 

If  some  unknown  error  appears  during  measuring,  remove  the  battery  to  terminate 

operating. 

 

Do not contort or drag the wire of the device. 

 

The  plethysmographic  waveform  is  not  normalized,  as  a  signal  inadequacy  indicator, 

when  it  is  not  smooth  and  stable,  the  accuracy  of  the  measured  value  may  degrade. 

When it tends to be smooth and stable, the measured value read is the optimal and the 

waveform at this time is also the most standard.   

 

.If  the  device  or  component  is  intended  for  single-use,  then  the  repeated  use  of  these 

parts  will  pose  risks  on  the  parameters  and  technical  parameters  of  the  equipment 

known to the manufacturer. 

 

If  necessary,  our  company  can  provide  some  information  (such  as  circuit  diagrams, 

component lists, illustrations, etc.), so that the qualified technical personnel of the user 

can repair the device components designated by our company.   

 

The  measured  results  will  be  influenced  by  the  external  colouring  agent  (such  as  nail 

polish, colouring agent or color skin care products, etc.), so don't use them on the test 

site. 

 

As to the fingers which are too cold or too thin or whose fingernail is too long, it may 

affect the measured results, so please insert the thicker finger such as thumb or middle 

finger deeply enough into the probe when measuring. 

 

The  finger  should  be  placed  correctly  (see  Attached  figure 5),  as improper  installation 

or improper contact position for sensor will influence the measurement. 

 

The  light  between  the  photoelectric  receiving  tube  and  the  light-emitting  tube  of  the 
device must pass through the subject’s arteriole. Make sure the optical path is free from 
any optical obstacles like rubberized fabric, to avoid inaccurate results. 

 

Excessive  ambient  light  may  affect  the  measured  results,  such  as  surgical  light 

(especially  xenon  light  sources),  bilirubin  lamp,  fluorescent  lamp,  infrared  heater  and 

direct  sunlight,  etc.  In  order  to  prevent  interference  from  ambient  light,  make  sure  to 

place the sensor properly and cover the sensor with opaque material. 

 

Frequent movement  (active  or  passive)  of  the  subject  or  severe  activity  can affect  the 

measured accuracy. 

 

The Pulse Oximeter should not be placed on a limb with the blood pressure cuff, arterial 

ductus or intraluminal tube. 

 

The measured value may be inaccurate during defibrillation and in a short period after 

defibrillation, as it has not defibrillation function. 

 

The device has been calibrated before leaving factory. 

 

The device is calibrated to display functional oxygen saturation.     

 

The  equipment  connected  with  the  Oximeter  interface  should  comply  with  the 

requirements of IEC 60601-1. 

1.4.2 Clinical restriction   

A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of 

subject  is  required.  For  a  subject  with  weak  pulse  due  to  shock,  low  ambient/body 

temperature,  major  bleeding,  or  use  of  vascular  contracting  drug,  the  SpO

2

  waveform 

(PLETH) will decrease. In this case, the measurement will be more sensitive to interference. 

B. The measurement will be influenced by intravascular staining agents (such as indocyanine 

green or methylene blue), skin pigmentation. 

C.  The  measured  value  may  be  normal  seemingly  for  the  tester  who  has  anemia  or 

dysfunctional  hemoglobin(such  as  carboxyhaemoglobin  (COHb),  methaemoglobin  (MetHb) 

and  sulfhaemoglobin  (SuHb)),  but  the  tester  may  appear  hypoxia,  it  is  recommended  to 

perform further assessment according the clinical situations and symptoms. 

D. Pulse oxygen only has a reference meaning for anemia and toxic hypoxia, as some severe 

anemia patients still show better pulse oxygen measured valued. 

E. Contraindication: no 

2 Principle 

 

Figure 1 Operating principle 

Principle of the Oximeter is as follows: An experience formula of data process is established 

taking  use  of  Lambert  Beer  Law  according  to  Spectrum  Absorption  Characteristics  of 

Reductive  Hemoglobin  (Hb)  and  Oxyhemoglobin  (HbO

2

)  in  glow  &  near-infrared  zones. 

Operation principle of the device is: Photoelectric Oxyhemoglobin Inspection Technology is 

adopted  in  accordance  with  Capacity  Pulse  Scanning  &  Recording Technology,  so  that two 

beams  of  different  wavelength  of  lights  can  be  focused  onto  human  nail  tip  through 

perspective  clamp  finger-type  sensor.  Then    measured  signal  can  be  obtained  by  a 

photosensitive element, information acquired through which will be shown on screen through 

treatment in electronic circuits and microprocessor.   

3 Functions 

A. SpO

2

 value display 

B. PR value and bar graph display 

C. Low-battery indication: low-battery indication appears when the battery voltage is 

too low to work 

D. Automatic standby function

 

                                                   

4 Installation 

4.1 View of the front panel 

 

Figure 2    Front View 

4.2 Battery 

Step 1. Refer to Figure 3-1 and insert the two AAA size batteries properly in the right 

direction. 

Step 2. Replace the cover.   

Please  take  care  when  you  insert  the  batteries  for  the  improper  insertion 

may damage the device.   

Sketch map for the back cover in ON/OFF state refer to Figure 3-2:

 

 

 

 

 

Figure 3-1 Batteries Installation        Figure 3-2 Installation for the Back Cover                 

 

4.3 Mounting the hanging rope   

Step 1. Put the end of the rope through the hole. 

Step 2. Put another end of the rope through the first one and then tighten it. 

 

Figure 4. Mounting the hanging rope. 

4.4 Structure, accessories and software description   

A. Structure: main unit. 

B. Accessories: one User Manual, one hanging rope.   

Please check the device and accessories according to the list to avoid that the 

device can not work normally.

   

C.

   

Software description   

Software name: CMS50DL embedded software 

Software specification: no 

Release version: V2.0 

Naming rule for version:  V <Major enhancive software upgrade>.<Minor enhancive 

software upgrade>.<Improvement software upgrade> 

Involved algorithm: name: plethysmography; type: mature arithmetic 

Purpose: be used to measure SpO

2

, pulse rate, etc. 

Clinical  function:  calculate  SpO

2

 and  pulse rate  values  by  collecting  and  processing 

the testee’ s pulse signal. 

5 Operating Guide 

5.1    Insert the two batteries properly to the direction, and then replace the 

cover. 

5.2    Open the clip as shown in Figure 5.   

 

Figure 5. Put finger in position 

5.3  Let the patient’s finger put into the rubber cushions of the clip (make sure 

the finger is in the right position), and then clip the finger. 

5.4  Press the button once on front panel. 

5.5  Do not shake the finger and keep the patient at ease during the 

process.Meanwhile, human body is not recommended in movement status. 

5.6  Get the information directly from screen display. 