EKG6012
32
Note: If the interference can not be removed after taking above measures, please use a
EMG filter. The ECG waveform recorded at this time will be slightly attenuated.
8.4 Baseline drift
➢
Whether the electrode installation is stable?
➢
Whether the connection of lead cables or electrodes is reliable?
➢
Whether the electrodes and patient skin are cleaned and are daubed with enough
conductive paste?
➢
Whether it is caused by patient's movement or breathing?
➢
Whether the electrodes or leads are in bad connection?
Note: If the interference can not be removed after taking above measures, please use a
baseline filter.
8.5 Troubleshooting list
Phenomenon
Cause of failure
Solutions
Too large
interference,
disorderly
waveform
1. Grounding cable is not
connected reliably.
2. Lead cables are not connected
reliably.
3. There is AC interference.
4. Patient is nervous and can not
keep quiet.
1. Check the power cord and
lead cables.
2. Let the patient prepare for the
measurement.
Baseline burr
1. AC interference is large.
2. Patient nervous, and EMG
interference is large.
1. Improve the environment.
2. If the bed is made of steel,
replace it.
3. The power cable and lead
cables are not parallel or too
close to each other.
Not
regular
1. Bad electrode conductivity.
1. Use alcohol of high quality.