background image

 

 

CMS50S     

User Manual

 

P u l s e   O x i m e t e r

                       

 

Contec Medical Systems Co., Ltd. 

Address:No.112 Qinhuang West Street, Economic 

 

Technical Development Zone,Qinhuangdao,Hebei   
Province, 

PEOPLE’S REPUBLIC OF 

CHINA 

Tel: +86-335-8015430 
Fax:+86-335-8015588   
Technical s86-335-8015431 
E-mail:[email protected] 
Website:http://www.contecmed.com 

EC REPRESENTATIVE 

Shanghai International Holding Corp. GmbH(Europe) 

Address: Eiffestrasse 80, 20537, Hamburg, Germany   

Tel: 

49-40-2513175   

Fax: 

49-40-255726 

E-mail

[email protected]

 

CMS2.782.492(CE)ESS/1.0            1.4.01.01.798                      2021.04 

User Notice 

Dear  users,  thank  you  very  much  for  purchasing  the  Pulse  Oximeter  (hereinafter 
referred to as device). 
This  Manual  is  written  and  compiled  in  accordance  with  the  council  directive 
MDD93/42/EEC  for  medical  devices  and  harmonized  standards.  In  case  of 
modifications and  software  upgrades, the  information  contained  in this document is 
subject to change without notice. 
It is a medical device, which can be used repeatedly. 
The  Manual  describes,  in  accordance  with  the  device

’s  features  and  requirements, 

main  structure,  functions,  specifications,  correct  methods  for  transportation, 
installation,  usage,  operation,  repair,  maintenance  and  storage,  etc.  as  well  as  the 
safety procedures to protect both the user and device. Refer to the respective chapters 
for details. 
Please  read  the  User  Manual  carefully  before  using  this  device.  The  User  Manual 
which  describes  the  operating  procedures  should  be  followed  strictly.  Failure  to 
follow  the  User  Manual  may  cause  measuring  abnormality,  device  damage  and 
human  injury.  The  manufacturer  is  NOT  responsible  for  the  safety,  reliability  and 
performance  issues  and  any  monitoring  abnormality,  human  injury  and  device 

damage  due  to  users'  negligence  of  the  operation  instructions.  The  manufacturer’s 

warranty service does not cover such faults. 
Owing to the forthcoming renovation, the specific products you received may not be 
totally  in accordance  with the  description  of  this  User  Manual.  We  would  sincerely 
regret for that. 
Our company has the final interpretation to this manual. The content of this manual is 
subject to change without prior notice. 

Warnings 
Remind that it may cause serious consequences to tester, user or environment. 

 

Explosive  hazard

DO  NOT  use  the  device  in  environment  with  inflammable 

gas such as anesthetic. 

 

DO NOT use the device while examining by MRI or CT, as the induced current 
may cause burn. 

 

Do  not  take  the  information  displayed  on  the  device  as  the  sole  basis  for  clinical 
diagnosis. The device is only used as an auxiliary means in diagnosis. And it must be 

used in conjunction with doctor’s advice, clinical manifestations and symptoms.

 

 

The  maintenance  to  the  device  or  replacement  of  the  batterycan  only  be 
performed  by  qualified  service  personnel  specified  by  manufacturer,  dangers 
(such  as  over-temperature,  fire  or  explosion)  may  occur  when  replacing  the 
battery by the personnel not fully trained. Users are not permitted to maintain or 
refit the device by themselves. 

 

Uncomfortable  or  painful  feeling  may  appear  if  using  the  device  ceaselessly, 
especially for the microcirculation disturbance users. It is not recommended that 
the sensor is used on the same finger for more than 2 hours. 

 

For  some  special  users  who  need  a  more  careful  inspection  on  the  test  site, 

please don’t place the device on the edema or tender tissue.

 

 

Please  do  not  stare  at  the  red  and  infrared  light  emitter  (the  infrared  light  is 
invisible) after turning on the device, including the maintenance staff, as it may 
be harmful to the eyes. 

 

The  device  contains  silicone,  PVC,  TPU,  TPE  and  ABS  materials,  whose 
biocompatibility  has  been  tested  in  accordance  with  the  requirements  in  ISO 
10993-1, and it has  passed  the recommended  biocompatibility  test.  The  person 
who is allergic to silicone, PVC, TPU, TPE or ABS can not use this device. 

 

The disposal of scrap device, its accessories and packaging should follow the local 
laws and regulations, to avoid polluting to the local environment. And the packaging 
materials must be placed in the region where the children are out of reaching. 

 

The device can not be used with the equipment not specified in the Manual. Only the 
accessories appointed or recommended by the manufacturer can be used, otherwise it 
may cause injury to the tester and operator or damage to the device. 

 

The  SpO

2

  probe  accompanied  is  only  suitable  for  using  with  the  device.  The 

device can only use the SpO

2

 probe described in the Manual, so the operator has 

the  responsibility  to  check  the  compatibility  between  the  device  and  the  SpO

2

 

probe  before  using,  incompatible  accessories  may  cause  device  performance 
degradation, device damage or user injury. 

 

Do not reprocess the accompanying SpO

2

 probe. 

 

Check  the  device  before  use  to  make  sure  that  there  is  no  visible  damage  that 

may affect user’s safety and device performance. When there is obvious damage, please 

replace the damaged parts before use. 

 

When the message ―Sensor Off‖ or ―Sensor Fault‖ appears, it indicates that the SpO

probe is disconnected or line fault occurs. Check the connection of the SpO

2

 probe and 

whether  there  is  damage  for  the  probe,  if  necessary,  please  replace  the  probe  to  avoid 
risks. The probe fault will not result in a safety hazard. 

 

Functional testers  can  not  be  used  to  assess  the  accuracy  of  the  SpO

2

  probe  and  Pulse 

Oximeter. 

 

Some  functional  testers  or patient  simulators  can  be  used  to  verify  whether  the  device 
works normally,  for  example,  INDEX-2LFE  Simulator  (software  version:  3.00),  please 
refer to the Manual for the detailed operation steps. 

 

Some  functional  testers  or  patient  simulators  can  measure  the  accuracy  of  the  device 
copied calibration curve, but they can not be used to evaluate the device accuracy. 

 

When using the device, please keep it away from the equipment which can generate strong 
electric field or strong magnetic field. Using the device in an inappropriate environment may 
cause interference to the surrounding radio equipment or affect its working. 

 

When storing the device, keep it away from children, pets and insects to avoid affecting its 
performance. 

 

Do not place the device in places exposed to direct sunlight, high temperature, humidity, dust, 
cotton wool or easy to splash water, to avoid affecting its performance. 

 

The measured accuracy will be affected by the interference of electrosurgical equipment. 

 

When several products are used on the same  people simultaneously, danger may occur 
which is arisen from the overlap of leakage current. 

 

CO poisoning will appear excessive estimation, so it is not recommended to use the device. 

 

This device is not intended for treatment. 

 

The intended operator of the device may be a user. 

 

Avoid maintaining the device during using. 

 

The  device  should  be  operated  by  medical  personnel  via  professional  training,or 
non-medical personnel who have been guided. 

1 Overview 

The oxygen saturation is the percentage of HbO

2

 in the total Hb in the blood, so-called the O

2

 

concentration in the blood, it is an important physiological parameter for the respiratory and 
circulatory  system.  A  number  of  diseases  related  to  respiratory  system  may  cause  the 
decrease  of  SpO

2

  in  the  blood,  furthermore,  some  other  causes  such  as  the  malfunction  of 

human  body's  self-adjustment,  damages  during  surgery,  and  the  injuries  caused  by  some 
medical checkup would also lead to the difficulty of oxygen supply in human body, and the 
corresponding symptoms would appear as a consequence, such as vertigo, impotence, vomit 
etc. Serious symptoms might bring danger to human's life. Therefore, prompt information  of 
users'  SpO

2

  is  of  great  help  for  the  doctor  to  discover  the  potential  danger,  and  is  of  great 

importance in the clinical medical field.   
Insert the finger when measuring, the device will directly record the SpO

2

 value measured, it 

has higher accuracy and repeatability. 

1.1 Features

   

A. Easy to use. 
B. Small in volume, light in weight, convenient to carry. 
C. Low power consumption. 

1.2 Indication for Use

 

The  Pulse  Oximeter  can  be  used  in  measuring  the  pulse  oxygen  saturation  and  pulse  rate 
through finger. The product is suitable for being used in family, oxygen bar, physical care in 
sports (It can be used before or after doing sports, and it is not recommended to use the device 
during the process of having sport) and etc. 

1.3 Environment requirements 

Storage Environment 
  a) Temperature: -40

~+60

 

  b) Relative humidity: 

≤95%

 

  c) Atmospheric pressure: 500hPa~1060hPa 
Operating Environment 
  a) Tempe5 

~+40

 

  b) Relative Humidity: 

90% 

  c) Atmospheric pressure: 700hPa~1060hPa   

1.4 Precautions 
1.4.1 Attention 
Point  out  conditions  or  practices  that  may  cause  damage  to  the  device  or  other 
properties.

 

 

Before using the device, make sure that it locates in normal working state and operating 
environment. 

 

In  order  to  get  a  more  accurate  measurement,  it  should  be  used  in  a  quiet  and 
comfortable environment. 

 

When  the  device  is  carried  from  cold  environment  to  warm  or  humid  environment, 
please do not use it immediately, wait four hours at least is recommended. 

 

If some unknown error appears during measuring, press button to reset it. 

 

If the device is splashed or coagulated by water, please stop operating. 

 

DO NOT operate the device with sharp things. 

 

High temperature, high pressure, gas sterilizing or immersion disinfection for the device 
is  not  permitted.  Refer  to  User  Manual  in  the  relative  chapter  (6.1)  for  cleaning  and 
disinfection. Please set the device to standby mode before cleaning and disinfecting. 

 

The device is suitable for children and adult. 

 

The device may not be suitable for all users, if you can't get a satisfactory result, please 
stop using it. 

 

During  measuring,  when  abnormal  conditions  appear,  please  pull  out  your  finger  and 
reinsert it to measure again. 

 

Data averaging and signal processing have a delay in the upgrade of SpO

2

 data values. 

When  the  data  update  period  is  less  than  30  seconds,  the  time  for  obtaining  dynamic 
average values will increase, which is arisen from signal degradation, low perfusion or 
other interference, it depends on the PR value. 

 

The device has 3-year service life, date of manufacture: see the label. 

 

The  device  does  not  provide  over-limit  alarm  function  for  SpO

2

  and  PR,  so  it  is 

inapplicable for using in the place where need such function.   

 

The maximum temperature at the SpO2 probe -tissue interface should be less than 41

 

which is measured by the temperature tester. 

 

Do not contort or drag the wire of the device. 

 

The  plethysmographic  waveform  is  not  normalized,  as  a  signal  inadequacy  indicator, 
when  it  is  not  smooth  and  stable,  the  accuracy  of  the  measured  value  may  degrade. 
When it tends to be smooth and stable, the measured value read is the optimal and the 
waveform at this time is also the most standard.   

 

The device can not be used during charging. 

 

If  necessary,  please  visit  our  official  website  to  get  the  information  about  SpO

2

  probe 

that can be used with this device. 

 

If  the  device  or  component  is  intended  for  single-use,  then  the  repeated  use  of  these 
parts will pose risks on the parameters and technical parameters of the equipment known 
to the manufacturer. 

 

If  necessary,  our  company  can  provide  some  information  (such  as  circuit  diagrams, 
component lists, illustrations, etc.), so that the qualified technical personnel of the user 
can repair the device components designated by our company.   

 

The  measured  results  will  be  influenced  by  the  external  colouring  agent  (such  as  nail 
polish,  colouring agent  or  color  skin  care  products,  etc.),  so  don't  use  them  on the  test 
site. 

 

As to the fingers which are too cold or too thin or whose fingernail is too long, it may 
affect the measured results, so please insert the thicker finger such as thumb or middle 
finger deeply enough into the probe when measuring. 

 

The finger should be placed correctly (see Attached figure 3), as improper installation or 
improper contact position for sensor will influence the measurement. 

 

The  light  between  the  photoelectric  receiving  tube  and  the  light-emitting  tube  of  the 

device must pass through the subject’s arteriole. Make sure the optical path is free from 

any optical obstacles like rubberized fabric, to avoid inaccurate results. 

 

Excessive  ambient  light  may  affect  the  measured  results,  such  as  surgical  light 
(especially  xenon  light  sources),  bilirubin  lamp,  fluorescent  lamp,  infrared  heater  and 
direct  sunlight,  etc.  In  order  to  prevent  interference  from  ambient  light,  make  sure  to 
place the sensor properly and cover the sensor with opaque material. 

 

Frequent  movement  (active  or  passive)  of  the  subject  or  severe  activity  can  affect  the 
measured accuracy. 

 

The  SpO

2

  probe  should  not  be  placed  on  a  limb  with  the  blood  pressure  cuff,  arterial 

ductus or intraluminal tube. 

 

The measured  value may  be  inaccurate during  defibrillation and  in a  short  period after 
defibrillation, as it has not defibrillation function. 

 

The device has been calibrated before leaving factory. 

 

The device is calibrated to display functional oxygen saturation.     

 

The  equipment  connected  with  the  Oximeter  interface  should  comply  with  the 
requirements of IEC 60601-

 

 

Please  select  medical  power  adapter  to  charge  it,  when  connecting  the  special  adapter 
with the socket, make sure there is no shelter near the socket and it is easy to plug and 
unplug, otherwise the power will not be cut off in time when necessary, causes damage. 

1.4.2 Clinical restriction   

A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of 
subject  is  required.  For  a  subject  with  weak  pulse  due  to  shock,  low  ambient/body 
temperature,  major  bleeding,  or  use  of  vascular  contracting  drug,  the  SpO

2

  waveform 

(PLETH) will decrease. In this case, the measurement will be more sensitive to interference. 
B. The measurement will be influenced by intravascular staining agents (such as indocyanine 
green or methylene blue), skin pigmentation. 
C. The measured value may be normal seemingly for the tester who has anemia or 

dysfunctional hemoglobin(such as carboxyhaemoglobin (COHb), methaemoglobin (MetHb) 
and sulfhaemoglobin (SuHb)), but the tester may appear hypoxia, it is recommended to 
perform further assessment according the clinical situations and symptoms. 

D. Pulse oxygen only has a reference meaning for anemia and toxic hypoxia, as some severe 

anemia users still show better pulse oxygen measured valued. 

E. Contraindication: no 

2 Principle   

 

 

 

Figure 1 Operating principle 

Principle of the Oximeter is as follows: An experience formula of data process is established 
taking  use  of  Lambert  Beer  Law  according  to  Spectrum  Absorption  Characteristics  of 
Reductive  Hemoglobin  (Hb)  and  Oxyhemoglobin  (HbO

2

)  in  glow  &  near-infrared  zones. 

Operation principle of the device is: Photoelectric Oxyhemoglobin Inspection Technology is 
adopted  in  accordance  with  Capacity  Pulse  Scanning  &  Recording Technology,  so  that two 
beams  of  different  wavelength  of  lights  can  be  focused  onto  human  nail  tip  through 
perspective  clamp  finger-type  sensor.  Then  measured  signal  can  be  obtained  by  a 
photosensitive element, information acquired through which will be shown on screen through 

treatment in electronic circuits and microprocessor. 

3 Functions   

a.

 

Low-battery indication: low-battery indication appears when the battery voltage is 
too low to work. 

b.

 

Automatic standby function. 

c.

 

Memory function. 

d.

 

The data measured can be upload to the terminal equipment by wireless mode. 

e.

 

Charging function. 

4 Installation   

4.1 Appearance 

 

Figure 2 Appearance 

USB interface: connect with USB cable or SpO

2

 probe 

Button: exit/enter the standby mode. 

4.2 Installation of SpO

2

 probe 

Open the USB plug of the device, then insert the SpO

2

 probe interface into the USB 

port of the device. 

4.3 Connection of USB cable 

Open the USB plug of the device, insert the micro end into the interface of the device, 
the other end to the power adapter. 

4.4 Structure and accessories and software description 

A.  Structure:  main  unit,  SpO

2

  probe,  USB  cable,  power  adapter  (optional)  and 

Bluetooth adapter (optional). 
B.  Accessories:  one  SpO

2

  probe,  one  USB  cable,  one  power  adapter  (optional),  one 

User Manual, Bluetooth adapter (optional).   
Please check the device and accessories according to the list to avoid that the device 
can not work normally. 
C.Software description   
Software name: CMS50S embedded software 
Software specification: no 
Release version: 2.0 
Naming rule for version: V <Major enhancive software upgrade>.<Minor enhancive 
software upgrade>.<Improvement software upgrade> 
Involved algorithm: name: plethysmography; type: mature arithmetic 
Purpose: be used to measure SpO

2

, pulse rate, etc. 

Clinical  function:  calculate  SpO

2

 and  pulse rate  values  by  collecting  and  processing 

the testee’ s pulse signal,

 

5 Operating   

5.1 Measurement 

A. Insert the finger into the probe as shown in Figure 3.

 

 

 

Figure 3 Sketch map for finger placement 

(The appearance of actual probe may be different with the one shown as Figure 3, 

please refer to the actual probe.) 

Note:  when  inserting  the  finger,  the  light  emitting  from  the  sensor  must  be 
directly irradiated to the side of the fingernail. 
Note: during measuring, do not shake the finger and keep quiet, not move. 

A.

 

Long press the Button to exit the standby mode. 

B.

 

The  device  measures  and  saves  the  data  automatically,  the  data  need  to  be 

uploaded to PC for checking.   

Note:  please  synchronize  the  time  with the  master  device  when  using  it  for  the 
first time, refer to chapter 5.5 for relative operations. 
5.2 Exit/enter standby mode 

a.

 

Under the standby state, long press the Button to exit from it. 

b.

 

Under non-memory state, long press the Button to enter the standby mode. 

c.

 

If the device has not stored data, it will automatically enter standby mode after 30 
s;  Button  operation,  Bluetooth  communication  and  device  charging  can  reset  the 
standby time. 

5.3 Insufficient storage time 

The device will prompt by LED when the memory space is full, memory space is not 
enough  to  store  8  hours  or  battery  power  is  not  enough  to  test  8  hours.  Refer  to 
section 5.7 for LED status. 

Reviews: