background image

6

Measuring Frequency Response 

Detailed below are two  processes that can be used to confirm the 

system properly reproduces all of the required frequencies equally in 

the bandwidth of 100 to 5KHz. A Test Signal Generator such as the 

Contacta TSG1 will be required to perform these tests. Both methods 

will yield accurate results. 

Method I  -  Sine wave signal 

Send at a minimum the following sine wave signals through the hearing 

loop system without adjusting any of the audio or loop level controls: 

100Hz, 250 Hz, 500 Hz, 1000 Hz, 2500 Hz, and 5000 Hz. Using the Flat 

Spectrum mode record the RMS level generated by each of these. To 

meet the IEC specification the levels should not vary by any more than 

±3dB.  

Method II  -  Pink noise signal 

Send a pink noise signal into the hearing loop system and select the 

Third Octave mode on the meter. It will initially start at a 1000Hz center 

frequency.  

Pressing button “B” will cycle through the frequencies of 100Hz, 

200Hz, 500Hz, 1000Hz, 2500Hz and 5000Hz. 

 

IEC 60118-4 Notes and Requirements 
The standard as revised notes that within the  frequencies of 100Hz to 

5000Hz the hearing loop system should equally reproduce all signals.  At 

a minimum the systems is to be tested at 100Hz, 1KHz and 5KHz. 

Program RMS/Peak Levels 

The second mode is used to set up the  hearing loop signal level using an 

A-Weighted filter. It is very difficult to set the field strength to an average 

of –12dBA as it will constantly vary based on the program used to set up 

the system. We will use the PEAK read-ing to confirm that  both our 

design and equipment meet the IEC specification. Pressing  button “B” 

resets the PEAK reading. 

How to test then adjust the  loop system level: 
Once the hearing loop system has been installed, play some bandwidth 

limited (100Hz to 5KHz) pink noise or 1KHz sine (use the driver 

manufacturers recommendation) though the loop system. Walk 

throughout the audience area holding your meter vertically. 

Note the readings and confirm that the A-RMS level does not vary by 

more than ± 3dBA. This lets you know that the perceived signal level in 

the hearing aids will be the same, no matter where the person is sitting. 

Next: using program audio similar to what is normally used in the facility, 

adjust the audio program level to a level just above “normal”. Now turn 

on the hearing loop system and adjust the drive level until a –3dBA to 

0dBA PEAK level is obtained in the center of the loop. Reset the  PEAK 

level and confirm this level averages 0 dBA across the  audience listening 

plane. 

IEC 60118-4 Notes and Requirements 
The revised standard states that the signal level across the loop area 

should not vary by more than ±3dB. Once that is confirmed, the signal 

level based on the building's normal program should peak at 0 dB as 

referenced to  a 400mA/m. This will confirm there is adequate loop 

current to produce both the peak and average (average is -12dB or 100ma/

m) signal levels for T-Coil equipped hearing aids.

Summary of Contents for IL-CONTACTA-FSM

Page 1: ...1 Field Strength Meter User Guide IL CONTACTA FSM ...

Page 2: ...op Systems be installed and certified to the new IEC 60118 4 standard When these standards are met the user experience will be equal wherever they have the opportunity to use their T coils Our design team interviewed many installers throughout the industry which resulted in the following design goals 1 Very accurate and easy to read 2 Simple straightforward operation 3 Based on the IEC 60118 4 spe...

Page 3: ...current mode and the next press advances to the next meter mode 1 Background Noise Test with Max indication 2 Signal Strength A weighted RMS Peak 3 Field Strength Flat RMS Max 4 Third Octave Levels 100 200 500 1K 2 5K 5KHz B Button functions In modes 1 3 pressing the B button reset the MAX or PEAK readings In mode 4 pressing the B button advances the center frequency of the third octave filters Se...

Page 4: ...m will be used holding the meter in a vertical position at the listening plane height The important reading will be the MAX reading However it is important to watch the RMS reading If the MAX reading exceeds 32dBA readings above 32dBA will have a lower negative number for example 30dBA means there is more background noise than 35dBA you will need to document the areas where those higher noise leve...

Page 5: ...vel does not vary by more than 3dBA This lets you know that the perceived signal level in the hearing aids will be the same no matter where the person is sitting Next using program audio similar to what is normally used in the facility adjust the audio program level to a level just above normal Now turn on the hearing loop system and adjust the drive level until a 3dBA to 0dBA PEAK level is obtain...

Page 6: ... 500 Hz 1000 Hz 2500 Hz and 5000 Hz Using the Flat Spectrum mode record the RMS level generated by each of these To meet the IEC specification the levels should not vary by any more than 3dB Method II Pink noise signal Send a pink noise signal into the hearing loop system and select the Third Octave mode on the meter It will initially start at a 1000Hz center frequency Pressing button B will cycle...

Page 7: ...level headphone output type headphone volume display units dB mG uT and power timeout 5 to 30 or none 7 Method II Pink noise signal Cont Record the RMS level readings for each of the frequencies As in method I if the level does not vary by more than 3dB the sys tem as installed will meet the IEC specification This method was requested by the field engineer so they could run the test with one instr...

Page 8: ... last mode on the FSM and displays an accurate level which can then be demonstrated to the building We advise a level between 10dBV and 0dBV To use this feature cycle through the various mode un til you see Line In Flat Response on the display Now connect up the XLR adapter which has an XLR female on one end and an 1 8 TRS connector on the other end If you wish to make up your own adapter use the ...

Page 9: ... 0dB 400mA m 3 0 1dB resolution for levels over 32dB 0 1 dB True RMS on all features Pickup Coil Length wise of meter s longer dimension and Parallel to meter face Position noted on Product Calibrated at 1 000 Hz sine to read 0 dB at 5 03 mG Flat 1dB from 50 Hz to 10 000 Hz Class 2 meter specified in IEC 61672 1 Single 9v Battery External Power Jack Output A Weighted or Flat selectable 16x2 LCD Ch...

Page 10: ...10 ...

Page 11: ...11 ...

Page 12: ...12 Local dealer UK ROW 44 0 1732 223900 sales contacta co uk US Canada 1 616 392 3400 info contactainc com www contacta co uk ...

Reviews: