Calibration Instruction
1 Calibration Instruction for Salwico ST200 and
Salwico ST600EX
5103026_Salwico ST200 & Salwico ST600EX_Calibration Instruction_I0_EN_2019_C
1
1 Calibration Instruction for Salwico ST200
and Salwico ST600EX
1.1 Preliminaries
This instructions applies to the following products:
Figure 1. Salwico ST200. Part No. 4470
Figure 2. Salwico ST600EX. Part No. 4460.
NOTE!
It is recommended to periodically carry out a gas response check on the gas detector
to ensure correct operation.
The ST200 and ST600EX detectors for 0-100% LEL Hydrocarbon can be
calibrated and the out-signal adjusted.
Hint!
Reading at 0% is 4.0 mA, and full range 100% LEL is 20 mA.
Tools
Tools required for this operation are: