DS1M12 - Quick Start Guide
© Connective Peripherals Pte Ltd
Clearance Number : CP#032
Document Reference Number: CP_000042
DS1M12_Quick_Start_Guide 1.1
Run the executable DS1M12_Setup.exe by double
clicking on the file and follow the on screen prompts to
install the application.
Application Installation (DataLogger)
The datalogger application file can be downloaded from:
https://www.connectiveperipherals.com/uploader/DDL1M
12_Setup.zip
Extract the files to a convenient location on your local
drive.
Run the executable DDL1M12_Setup.exe by double
clicking on the file and follow the on screen prompts to
install the application.
The applications can then be accessed from the START>All
Programs list of applications.
Measuring High Voltages
The DS1M12 is rated at ± 50V DC maximum input (35V
RMS AC) even with a x10 oscilloscope lead attached.
Measuring higher voltages than this directly may damage
the instrument and invalidate the warranty. The instrument
is not galvanically isolated. The ground of the internal
circuitry connects directly to the ground of the host PC via
the USB connection. To measure high voltages such as
mains supplies, use a high voltage differential probe such as
the Connective Peripherals SI-9001 for your personal safety
and to prevent damage to the instrument.