15
3.3
Power Cord Requirements
100/120 Volt (Linvatec Cat. No. C7104)
Use only a listed (UL, CSA) detachable power cord manufactured to the following specifications.
•
Plug End
•
NEMA 5-15P hospital grade, 15 amps, 125V
•
Receptacle End
•
IEC 320/CEE-22, 6 amps, 250V/15 amps, 125V
•
Cord
•
UL style SJT, 14 AWG, 3 conductor
220/240 Volt (Linvatec Cat. No. C7105)
•
Plug End
•
Molded straight PVC plug with double grounding system
•
DIN 49441, CEE 7/U11, 10/16A, 250V
•
CEBEC, DEMKO, KEMA, NEMKO, OVE, SEMKO, VDE, UTE, FEMKO
•
Receptacle End
•
Molded straight PVC plug
•
DIN 49457, CEE 22/V, 10A, 250V
•
VDE, D, N, S, SEV, OVE, KEMA
•
Cord
•
PVC, 7.2mm diameter
•
10A, 250V
•
Conductors: 3 x 1 mm
2
•
Conductor Colors - brown, blue, green/yellow stripe
NOTE:
If the product is connected to a 240V system in the United States, then it must be
connected to a center tap system.
There are no toxic components contained in the Xenon Light Source. After the useful life of the
product, dispose of components and service parts properly.
Summary of Contents for Linvatec LS7500
Page 1: ...The Linvatec LS7500 300 Watt Xenon Light Source Instruction Manual...
Page 21: ...17...
Page 22: ...18...
Page 23: ......