CLEANING AND STERILIZING
2.1
Cleaning and Washer/Sanitizer Warnings and
Precautions
1. Follow universal precautions for protective apparel when handling and
cleaning contaminated instruments.
2. Clean instruments within 30 minutes after use to minimize the potential of
blood and debris drying prior to cleaning.
3. After use, shaver blades and burs may be a potential biohazard and
should be handled in accordance with acceptable medical practice and
applicable local and national requirements.
4. Do not lubricate handpieces.
5. Never immerse the handpieces. Corrosion to handpiece may
occur resulting in decreased handpiece performance.
6. For aluminum surfaces, a neutral-pH agent should be used. To prevent
corrosion, avoid contact with strong alkaline solutions (pH over 10.5) or agents
containing iodine or chlorine.
7. Prior to using a washer / sanitizer, consult product labeling on all washer / sanitizer
cleaning solutions for compatibility with aluminum. Use of a washer / sanitizer
may decrease the life expectancy of the handpiece.
8. Never clean handpieces with bleach, chlorine-based detergents, liquid or
chemical disinfectants, or any products containing sodium hydroxide (such as,
INSTRU-KLENZ or Buell Cleaner). These products degrade the anodized
aluminum coating and may result in reduced handpiece reliability.
9. Never clean equipment in an ultrasonic cleaner.