1-1

AER DEFENSE

TM

S M O K E E V A C U A T O R

General Information

Section 1.0

1.1 Introduction 

Indications for use: The CONMED AER 

DEFENSE™ Smoke Evacuator is designed to 

remove smoke, aerosols and mitigate odors 

produced by surgical smoke during electrosurgical 

procedures. 

The CONMED AER DEFENSE™ Smoke 

Evacuator has been designed with a high suction, 

high flow rate, thru flow vacuum motor. Its quiet 

motor is used to draw the surgical smoke from 

the surgical site, through the vacuum tubing and 

into the CONMED AER DEFENSE™ Smoke 

Evacuator’s FilterOne™ where the surgical smoke 

is processed by a series of filtration stages. A 

single disposable filter is used to simplify the 

installation and removal during filter changes. 

The FilterOne™ is completely enclosed to 

protect the health care personnel from potential 

contamination during filter changes. Each 

FilterOne™ contains three integral filter stages to 

capture the smoke plume. 

The first stage of filtration is a pre-filter whose 

function is to trap and remove gross particulates 

and casual fluid. 

The second stage of filtration uses the highest 

grade of virgin grade activated carbon for the 

removal and adsorption of odors and toxic gases 

produced by electrosurgical tissue effects. These 

harmful gases may constitute a health hazard 

to health care professionals who are subjected 

to prolonged exposure. The activated carbon 

used in the FilterOne™ preferentially removes 

toxic organic gases rather than water vapor and 

provides optimal odor removal. 

The third stage of filtration is a ULPA (Ultra 

Low Penetration Air) grade filter whose high-tech 

design captures particulates and micro-organisms 

down to 0.10 µm at an efficiency of 99.9995%. 

Each ULPA grade media filter is certified  

before use. 

The electronic controls on the front panel of the 

CONMED AER DEFENSE™ Smoke Evacuator 

have been designed to be “user friendly” and 

facilitate unit set-up and operation. Please refer to 

Section 2.0 for operating instructions. 

1.2 Inspection 

The CONMED AER DEFENSE™ Smoke 

Evacuator has been thoroughly tested and 

inspected before shipment from the factory. Please 

check the unit before using it to ensure that no 

damage has occurred in transit. If damage is 

evident, please contact CONMED Customer 

Service at one of the numbers listed on the inside 

front cover of this manual. 

The 100-120 VAC version of the CONMED 

AER DEFENSE™ Smoke Evacuator includes an 

AC power cord for connection to a standard (US) 

outlet, 4’ ESU Activation Cord and an Operator’s 

Manual. The 4’ ESU Activation Cord is shipped 

with the 220-240 VAC version of the smoke 

evacuator; the AC power cord and Operator’s 

Manual are not. 

Operator’s Manuals in languages other than 

English and mains power cords for use in other 

countries are available from CONMED. 

Please contact CONMED Customer Service at the 

numbers listed on the inside front cover of this 

manual to order approved accessories available 

for use with the CONMED AER DEFENSE™ 

Smoke Evacuator. 

CAUTION:

  FEDERAL LAW (USA) RESTRICTS THIS DEVICE TO SALE BY OR ON THE 

ORDER OF A PHYSICIAN.

Summary of Contents for AER DEFENSE

Page 1: ...Operator s Manual C M ON ED E L E C T R O S U R G E R Y TM AER DEFENSETM S M O K E E V A C U A T O R...

Page 2: ...tica NY 13502 5994 USA Patent Pending For Technical Service or Return Authorization Phone 303 699 7600 1 800 552 0138 Extension 5274 Fax 303 699 1628 For Customer Service or to order parts phone 1 800...

Page 3: ...1 5 1 5 2 Rear Panel 1 6 2 0 Operation 2 1 2 1 System Controls 2 1 2 1 1 Control Panel 2 2 2 1 2 Rear Panel 2 3 2 2 FilterOne Installation 2 4 2 3 Set up and Operation 2 5 2 3 1 AER DEFENSE ESU Activa...

Page 4: ...AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Page 5: ...es toxic organic gases rather than water vapor and provides optimal odor removal The third stage of filtration is a ULPA Ultra Low Penetration Air grade filter whose high tech design captures particul...

Page 6: ...ded receptacle of the same voltage as listed on the nameplate on the smoke evacuator WARNING To avoid risk of electric shock this equipment must only be connected to a supply mains with protective ear...

Page 7: ...will void the warranty If filter is inserted into the AER DEFENSE and the filter is not recognize and no green LED lights illuminate check to make sure filter is inserted properly and completely If fi...

Page 8: ...s device to ensure compatibility with other electrical devices List of compatible CONMED accessories 60 8054 001 Adapter non sterile 60 6828 005 ClearVac 60 6829 005 ClearVac 60 6828 001 ClearVac 60 6...

Page 9: ...ter 1 5 Explanation of Symbols 1 5 1 Control Panel 1 4 1 Accessory Diagram All adapters tubing ClearVac and GoldVac smoke pencils attach to the front nozzle on the FilterOne Image Shown 60 7510 005 Go...

Page 10: ...icing to qualified service personnel Refer to Instruction Manual Booklet for critical safety instruction The color of the symbol is blue 9 Defibrillation Proof Type CF Applied part 9083 Unit functions...

Page 11: ...s default mode when none of the available specialty modes have been selected and illuminated on the control panel Laparoscopic Continuous Automatic or Footswitch While in Normal Mode the AER DEFENSE S...

Page 12: ...SU Automatic Mode Laparoscopic Mode indication of the status of the life of the filter in use Each LED segment represents 1 8 or 12 5 of the filter life The life of the filter is a minimum of 10 hours...

Page 13: ...e 2 1 2 Rear Panel T10 0A H 250V REPLACE FUSES AS MARKED 100 120V 50 60Hz 12A 120V 6 3A ESU Activation Outlet RS232 Connector AER DEFENSE Fuses AER DEFENSE Mains Power Inlet Pneumatic Footswitch Conne...

Page 14: ...e the unit If the AER DEFENSE is turned back on the Orange LED in the filter life indicator will flash and there is an audible warning 5 beeps to indicate that a new filter must be installed The unit...

Page 15: ...ange this filter may result in decreased efficiency and contamination of the electric motor vacuum pump and sound absorbing media within the unit 2 3 Set up and Operation with Instant ESU Activation T...

Page 16: ...or down arrow buttons on the control panel while the unit is activated Noise created by the smoke evacuator may be minimized by selecting the lowest suction setting that effectively clears the sterile...

Page 17: ...O K E E V A C U A T O R Figure 2 3 1 AER DEFENSE ESU Activation Connection Set Up CORD TO ESU ESU ACTIVATION AER DEFENSE Power Cord ESU Activation Cord T10 0A H 250V REPLACE FUSES AS MARKED 100 120V...

Page 18: ...2 8 AER DEFENSETM S M O K E E V A C U A T O R AER DEFENSE Power Cord ESU Activation Cord...

Page 19: ...th 8 2 20 8 cm Length 18 7 47 5 cm Weight 20 lbs 9 07 kg Filter ULPA grade with Carbon Filtration Efficiency 99 999 0 12 m particles Tubing Sizes 7 8 22 mm 6 4 mm Operating Temperature 10 C to 30 C Tr...

Page 20: ...2 10 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Page 21: ...Troubleshooting the AER DEFENSE Problem Potential Cause s Corrective Actions Smoke Evacuation System is on but suction is minimal 1 FilterOne not seated completely 2 FilterOne clogged 3 Vacuum hose cl...

Page 22: ...3 2 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...

Page 23: ...cover of this manual Note the R A number on the outside of the carton 4 2 Service Ordering Parts To reorder obtain replacement parts or to return a unit for service call Customer Service at the number...

Page 24: ...4 2 AER DEFENSETM S M O K E E V A C U A T O R This page intentionally left blank...