Inflow/Outflow/Shaver Suction Tubing Set Installation

Choose Surgery Preset

 With pronged side of the inflow 

tubing cassette facing the 

console, position the  tubing 

loop over the large blue rotor. 

Pull cassette down so that the 

cassette snaps into place.

This button is used to select preset surgical settings. 

Pressing this button toggles between the preset selections 

for knee, shoulder, small joint, and hip surgical 

procedures. Each preset adjusts the pressure, outflow 

and shaver suction for the specified procedure.  

Note

: The Knee preset also allows the pump to 

operate in Wave Mode designed for optimum flow  

in knee surgery. 
Press the Run/Stop button to 

start and stop the flow of the 

pump. When the green LED  

is lighted the pump is in the  

Run position.

With pronged side of outflow/

suction tubing cassette facing 

the console, position the tubing 

loop over the small silver rotor. 

Pull the cassette down so that 

the cassette snaps into place.

Push on the cassette tab to 

insure that it is latched.

» Close the clamps on the bag spike line.
» Spike the fluid bags.
»  To prime the tube set, press the Run/Stop button 

until the air is purged from the line, press the 

Run/Stop button again to stop the pump.

»  Connect the patient ends of the tube sets to the 

proper instruments.

Knee/Wave Mode

Shoulder

Small Joint

Hip

1

2

Intended Use 

The ConMed Linvatec 24k

®

 Irrigation System is intended to provide fluid distention, irrigation, and suction during diagnostic and operative 

knee, shoulder, hip, and small bone arthroscopic procedures, and fluid irrigation during laparoscopic procedures.

Please review the 24k Instructions for Use Manual for complete operating instructions.