24k
Inflow/Outflow/Shaver Suction Tubing Set Installation
Choose Surgery Preset
With pronged side of the inflow
tubing cassette facing the
console, position the tubing
loop over the large blue rotor.
Pull cassette down so that the
cassette snaps into place.
This button is used to select preset surgical settings.
Pressing this button toggles between the preset selections
for knee, shoulder, small joint, and hip surgical
procedures. Each preset adjusts the pressure, outflow
and shaver suction for the specified procedure.
Note
: The Knee preset also allows the pump to
operate in Wave Mode designed for optimum flow
in knee surgery.
Press the Run/Stop button to
start and stop the flow of the
pump. When the green LED
is lighted the pump is in the
Run position.
With pronged side of outflow/
suction tubing cassette facing
the console, position the tubing
loop over the small silver rotor.
Pull the cassette down so that
the cassette snaps into place.
Push on the cassette tab to
insure that it is latched.
» Close the clamps on the bag spike line.
» Spike the fluid bags.
» To prime the tube set, press the Run/Stop button
until the air is purged from the line, press the
Run/Stop button again to stop the pump.
» Connect the patient ends of the tube sets to the
proper instruments.
Knee/Wave Mode
Shoulder
Small Joint
Hip
1
2
Intended Use
The ConMed Linvatec 24k
®
Irrigation System is intended to provide fluid distention, irrigation, and suction during diagnostic and operative
knee, shoulder, hip, and small bone arthroscopic procedures, and fluid irrigation during laparoscopic procedures.
Please review the 24k Instructions for Use Manual for complete operating instructions.