ATEMP/DTEMP SETTING MODE
1. Press “Max/Min/Avg” key again after Emissivity mode, it reaches the
Atemp/Dtemp setting mode.
2. Press Activation Trigger to change the display of Atemp or Dtemp.
3. Press “Max/Min/Avg” key or wait for exit from setting mode to confirm
the selection.
Celsius and Fahrenheit
1. Users can change the temperature unit from Celsius to Fahrenheit and
vise versa by sliding the C/F switch.
LOW BATTERY
1. If the unit is low in battery power,
the LCD flashes the battery icon on
the top left corner.
2. If the unit is almost out of power,
the “BAT” will show on the LCD
display.
Flashing
Toggle with trigger button
DEW POINT LEVEL
1. The bar on the left indicates the degree of occurrence of fog or dew on
surface material being scanned. E.g. if the surface temperature is below that
of the Dew Point, the bar shows full segments showing fog or dew can occur.
2. If the surface temperature is 2.5 degrees below Dew Point, the bar on the
left displays 10 segments.
3. If the surface temperature is same as Dew Point, the bar on the left
displays 5 segment.
4. If the surface temperature is 2.5 degrees above Dew Point, the bar is left
blank showing no fog or dew occurs.
Surface temperature
is 2.5 degree
below Dew point
Surface temperature
is same as Dew point
Surface temperature
is 2.5 degree
above Dew point