User's Guide to the PON-3

 

 

PON Power Meter

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

PON-3

 

 

English Version

 

Summary of Contents for PON-3

Page 1: ...User s Guide to the PON 3 PON Power Meter PON 3 English Version...

Page 2: ...User s Guide to the PON 3 PON Power Meter...

Page 3: ...reparing for Operation 5 Operation 5 1 Layout 5 2 Key s Functions 5 3 Definitions 5 4 Quick Operation 5 5 Detail Operation 5 5 1 Powering On the Instrument 5 5 2 Test mode switch 5 5 3 Threshold Setup...

Page 4: ...90nm and 1550nm that carry voice data and video in a single fiber The PON 3 can measure not only 1490nm and 1550nm optical signal but also accurately detect and measure the upstream burst at 1310nm se...

Page 5: ...D indicators for easy visibility Simply connect and display the results for all three wavelengths 1310 1490 1550nm Detect and measure the1310 nm upstream burst Supports APON BPON EPON and GPON network...

Page 6: ...m Insertion Loss 1 5dB Accuracy 1 0 5dB Linearity 0 2dB Data Storage 1000Sets Display 2 8 inch TFT LCD Refresh Rate of Display 2 5Hz Threshold 10 sets configured via PC based software Auto Power off Y...

Page 7: ...date of delivery to the original customer Any product found to be defective within the warranty period can be returned to a ComSonics authorized service center for repair replacement and calibration...

Page 8: ...electric shock or equipment damage Connect AC power cord with the equipment and wall socket properly While inserting the AC plug make sure there is no dust or dirt on the terminals and both plugs are...

Page 9: ...AC adapter 2 Please make sure that you have turned the instrument OFF before charging the batteries unplug the AC adapter when the batteries are fully charged 3 Please make sure the batteries are cor...

Page 10: ...input jack on the left side of the PON 3 device When the AC adapter is plugged in the AC Indicator on the LCD will be displayed Note 1 Power is supplied by the AC adapter even if the batteries are ins...

Page 11: ...strument during transit Checking the package contents The standard accessories of PON 3 are as follows PON 3 Device FC and SC Adaptors Calibration Certificate 4 Ni MH Batteries Carrying Case AC Charge...

Page 12: ...o parts Part I LCD Display Part II Key Matrix and LED Indicators The device shows the test results on the LCD screen as well as the status by the LED s Device connector types FC PC SC PC ST PC optiona...

Page 13: ...ve current data Menu mode Left arrow MENU Enter menu mode press again to exit MODE Activate Deactivate Pass Fail mode including LED indicators Menu Mode Up Arrow dBm dB Switches between dBm and dB mea...

Page 14: ...Warning GREEN Pass Thresholds for Pass Fail The definitions for the three states Fail Warning and Pass are described as below There are 5 Threshold limits Limit1 PON 3 upper measurement capability of...

Page 15: ...key to Select the option Select the Threshold setting that applies to your application using the same keys Then press MENU to return 4 The device will automatically make test measurements and display...

Page 16: ...ON and is shown in the bottom of the LCD screen If you want to turn it off press the Power key and release quickly less than 2 seconds to deactivate Auto shut down function The auto shut down functio...

Page 17: ...m key to switch the units See Fig 1 10 03 03 11 43 15 Auto off dBm units display the absolute power level dB units display a power level relative to a Reference value In this mode a reference value ne...

Page 18: ...under the resources download section In the Threshold menu view the top line indicates system information date time and the bottom line indicates the threshold information including threshold number...

Page 19: ...r When the cursor moves to a number the user can use key to increase value and to decrease value When cursor moves to Yes press the REF SEL key for more than 2 seconds the Yes will flash which means t...

Page 20: ...etween dBm and dB The dBm unit is the actual absolute power and dB unit is a power value relative to reference value Refer to next section about Reference value setup When the instrument is in the Pas...

Page 21: ...values The test results after setting the reference are relative to the reference value which is useful to measure deltas The units will be displayed in dB and the LED indicators will not light In the...

Page 22: ...7 10 03 03 11 43 15 Fig 7 Backlit Setup Menu In this menu press the Mode key to brighten the backlight and the dBm dB key to lower the backlight After completing the setup press REF SEL key less than...

Page 23: ...the example shown on the right the record number is num 0024 There is a 1 0000 record maximum that can be stored Press the SAVE button for longer than 3 seconds to save the current test measurements...

Page 24: ...ves Skip to the previous 100 memory saves Skip to the next 100 memory saves Exit out of the record view back to the main menu X Deletes the record with the X next to it on the right hand side RST Rese...

Page 25: ...ge the instrument battery 2 While charging the charging indicator will be active on the LCD screen to show the battery charge status 3 ONLY use rechargeable batteries to prevent damage when the AC DC...

Page 26: ...e device is not in use 2 Keep the optical connectors away from oil dirt and other contamination to ensure the proper operation Clean the connectors on a regular basis to ensure they are free from cont...