Compur Statox 4120
10. Technical data
General:
Gas Measuring
range
Cl
2
0-1.5 ppm, 0-3 ppm, 0-10 ppm
CO 0-150
ppm
COCl
2
0-0.1 ppm, 0-0.3 ppm, 0-0.5 ppm, 0-1.5 ppm, 0-15 ppm, 0-100 ppm
HCN
0-15 ppm, 0-30 ppm, 0-50 ppm, 0-100 ppm
HCl 0-100
ppm
H
2
S
0-30 ppm, 0-50 ppm, 0-100 ppm
ClO
2
0-0.5
ppm
NO
2
0-15
ppm
SO
2
0-5
ppm
Accuracy at TLV:
+/- 10 %
Alarms:
2 adjustable alarm thresholds, system alarm
RFI: CE
conform
Manufacturer:
Compur Monitors, Munich
Sensor head:
Type:
5330 xxx (xxx: measuring range / gas)
Weight: 1.9
kg
Dimensions (mm, HxBxT):
180x300x180
Operating temperature:
-20 to +40
o
C
Storage temperature:
-30 to +50
o
C
Humidity:
20-95 % r.H. (not condensing)
Pressure: 800-1200
hPa
Protection class:
IP53
Approvals (div.countries):
Cl
2
, COCl
2
, H
2
S
Explosion protection:
EEx ib IIC T6
Operating environment :
II 2 G
Operating voltage U
i
:
max. 22 VDC
Operating current I
i
:
max. 50 mADC
Internal capacity C
i
: 55
nF
Internal inductivity L
i
: 0
mH
30