G.  Operation of the Scale (weighing powder):

1.  Turn the scale on by pressing the "ON" key and wait until the display reads "0.00" 

grams.  If the scale is not in the mode desired, then press the MODE key to change the 

weight unit accordingly.  The selected weight capacity (grams or grains for example) 

will remain even when the scale is turned off and then on again.  Once the proper weight 

capacity is selected, the scale is ready to weigh powder.

2   Place an empty powder pan on the scale.  

FIG 12

3.  If there are environmental variables affecting the scale's performance, then lower the 

Draft Shield to stabilize the readings.

4.   Press the ZERO key to 

"zero out" the scale.  This 

will remove the powder 

pan weight allowing only 

the powder itself to be 

weighed.  This function is 

called "TARE Function"  

FIG 11

5.   Wait till the scale reads 

"0.00".  At this point the 

scale can be used to weigh 

any object that will fit on 

the platform (within the 

weight range of the scale).

6.   Pour powder into the 

powder pan.  Allow 3-4 

seconds for the display 

to settle (because of the 

scale's extreme sensitivity, 

the display may fluctuate 

slightly, if so, check for 

vibration or air move-

ment).  Read the weight of 

the scale.

7.   Turn the scale off when 

you are finished.

FIG 12

F.  Re- Zero Function:

Press the "ZERO" key at any time to 

re-zero the scale.  This is a good idea if it 

is moved during a weighing process. 

FIG 11

FIG 11

CED Professional Digital Scale manual.indd   5

14/10/2016   16:40

Summary of Contents for CED PRO II

Page 1: ...CED PRO II High Precision Electronic Scale Instructions Phone USA 610 610 366 9752 888 628 3233 USA Fax 610 366 9680 Hong Kong 852 28666802 CED Professional Digital Scale manual indd 1 14 10 2016 16 40 ...

Page 2: ...ti purpose usage The maximum capacity is 100 GRAMS or 1500 GRAINS and the resolution is 0 01 GRAMS or 0 1 GRAINS Features include an Anti Static draft shield Calibration function Tare function an Overload cell protector Automatic shutoff and universal AC or battery operation for portability and field use No other scale on the market today provides as many quality features and value as the CED PRO ...

Page 3: ...te readings due to environmental influences This specially designed cover incorporates an anti static material to eliminate fluctuations in readings This prevents inaccurate readings due to static electrical build up A feature normally found only on very expensive laboratory scales When used in indoor environments without influences the cover should be lifted up to allow easy access and use FIG 5 ...

Page 4: ...king on and off FIG 7 Press the ON OFF key once and the display will show CAL O which indicates that the zero point is set FIG 8 4 GENTLY place the 100 GRAM cali bration weight on the platform and wait until CAL F is displayed FIG 9 The scale is now calibrated and ready to use Remove the weight and wait for 0 00 to appear If 0 00 does not appear press ZERO to return the scale to zero If you wish t...

Page 5: ...ess the ZERO key to zero out the scale This will remove the powder pan weight allowing only the powder itself to be weighed This function is called TARE Function FIG 11 5 Wait till the scale reads 0 00 At this point the scale can be used to weigh any object that will fit on the platform within the weight range of the scale 6 Pour powder into the powder pan Allow 3 4 seconds for the display to sett...

Page 6: ...ital scales have lock in mechanisms designed to stabilize the readings and prevent constant fluctuations If the trickling function is performed TOO SLOWLY it is possible to fool the scale In this case the scale will remain locked on the reading displayed even though more powder is actually being added When this occurs it is possible to have a powder charge indicate a lighter weight than is actuali...

Page 7: ...ght 6 The scale should be kept clean and away from moisture Only clean with a clean dry cloth Technical Support USA 1 888 623 3233 Hong Kong 852 28666802 The CED PRO II digital scale is covered by a one year manufacturer s warranty PACKAGE CONTENTS 1 CED PRO II Scale 4 AAAlkaline Batteries 1 AC Adapter 1 Attached Weighing Platform 1 100 GRAM Calibration Weight 1 Powder Pan 1 Attached Draft Shield ...

Page 8: ...This page can add you others products CED Professional Digital Scale manual indd 8 14 10 2016 16 40 ...

Reviews: