TECHNICAL SPECIFICATIONS
Electrical Requirements
AC120V / 60 Hz
Rated Current Input
3.5A
Sound Level
≤50.5 dBA
Altitude
Up to 6000’ without degradation of
concentration levels
Atmospheric Pressure Range: 12.47 psi-15.37
psi (86 kPa-106 kPa)
Oxygen Concentration
93% (±3%) at 0.5~5L/min after warmup
period of 30 min.
Maximum Liter Flow
5L/min
Maximum Outlet Pressure
5.5 psi ±0.7 psi (35 kPa ± 5 kPa)
Low Flow Alarm
0.5-5 L/min
Filters
Cabinet, Compressor Inlet, Final Bacteria
Product Dimensions
13”L x 10.25”W x 21.25”H
Product Weight
35.5 lbs
Operating Ambient Temperature and
Humidity
41°F-104°F (5°C-40°C),
≤80% RH
Storage and Transport Temperature
and Humidity
32°F-131°F (0°C-55°C)
10%-90% RH
Electrical Classification
Class II equipment, Type BF applied part,
IP21
Mode of Operation
Continuous
Minimum Operation Time
30 minutes
Safety System
Low Pressure Alarm
Low Oxygen Concentration Alarm
High Pressure/High Temperature Shutdown
Current Overload/Line Surge Shutdown
Maximum Cannula/Tubing Length
7’ C 50’ Oxygen Supply Tubing
Expected Life
6 Years
Warranty Period
3 Year Limited, 12,000 Hours on Parts
Software Version
A1.0
Flow Range at Outlet Pressure of 0 psi 0.5-5L/min
Flow Range at Outlet Pressure of 1 psi 0.5-5L/min
Change in maximum recommended
flow when back pressure of 1 psi is
applied
<0.5L/min
CAUTION: Technical specifications for
flowmeter calibration by manufacturer and
tester. This device
MUST
be used according
to manufacturer instructions and physician’s
order.
DO NOT
adjust the device flow.