background image

 

Measurement range  
Measurement modes

  - rpm & rps optically (also Count & Time)     

- rpm & rps, metres, yards, feet, per min & per sec.  via contact adaptor 
- Count total revs, metres, feet, yards 
- Measure Time interval in seconds between pulses (reciprocal rate). 
- Speed Capture feature - Maximum, Minimum or Average rate  

Speed range - Optical mode

 - 3 - 99,999 rpm   (or equivalent in rps)  

      

Contact mode  

- Max. 50,000rpm  for 10 sec (or equivalent in rps) 

 

Linear speeds -maximum 

- 0.3 - 1500.0 Metres or Yds/min. (4,500 ft/min) or equivalent in seconds 

Resolution range features 

- Fully Autoranging, up to 0.001 digit or +/- 1 digit fixed, User selectable  

Accuracy speed modes only  - 0.01% +/- 1 digit 
Count mode resolution 

- +/- 0.1 Metres (or equivalent in all ranges) 

Time interval mode   

- 0 - 99999 seconds Autoranging only  (max.  0.001 resolution) 

Timebase standard   

- 0.8 seconds or time between pulses, whichever is longest 

Timebase, Fast mode 

- 0.1 seconds auto-selection in Maximum or Minimum capture mode 

Memory features

  

- Last reading held for 1 minute, Auto Switch Off 
- Programme settings retained in memory after power down off

 

Contact adaptor 

 

- Included complete with rpm cone & metric wheel assembly (removable) 

Power requirements 

4 x AAA alkaline cells  

 

Standard package  
 

Contact adaptor, pack of reflective tape, connecting lead, certificate of calibration, 

 custom carrycase & instructions

Operation of model A2013/LSR 

 

Ex works setting - rpm mode, non autoranging 
1. Programming  - measurement mode selection 

All measurement modes are chosen by this method and once confirmed, selected mode remains in 
permanent memory until re-programmed by the user. 
1.1. 

To change mode hold programme button on and press up measure button and then release 

 

both buttons, the display will now illuminate all icons, and the current range will flash. 

1.2.  

To select new measurement mode press either up or down button to scan through the 
modes, when the   required mode icons flash release measure button & press programme 
button once to confirm settings. 

 

For non-speed modes the unit is now programmed and ready for use. 

1.3.  

To select 

mx, mn, av.

 modes continue to scan through each one, if the mode is not required, 

stop scan when all three icons illuminate continuously, then press programme button once. 

 

The instrument is now ready to use. 

Set parameters will be retained until reprogrammed 

2.  Optical revolutions speed measurement - rpm or rps.  

 

 

2.1. 

Ensure batteries are correctly fitted. 

2.2. 

Attach small reflective target to machine shaft (typically 6mm x 25mm, less for laser version)  

2.3. 

Start machine and point the tachometer towards the target. 

2.4. 

Press and hold either of the on/off buttons to suit application and hold continuously. 

2.5. 

Aim light beam onto target, ensure “on-target” sign is glowing or flashing steadily 

2.6. 

Read off rpm, releasing button will hold last reading.  

2.7. 

Last rpm reading will be held in display for 1 minute. 

2.8. 

Press the On button to zero reading or take another measurement.  

3.  Contact revolution speed measurement - rpm or rps.

   

 

 

3.1. 

Fit contact adapter into the tachometer and ensure a good click fit connection. 

3.2. 

Start machine and make clean contact with the recess in shaft end (wheel can be removed). 

3.3. 

Contact the shaft end via the rubber cone, ensure a steady firm pressure is applied and that 
the instrument is accurately in line with the machine shaft. 

3.4. 

Press and hold either up or down measure button as required & read speed 

3.5. 

Releasing the 

On

 switch will hold the reading automatically for 1 minute, automatic switch off. 

4.  Linear contact speeds measurement - metres, yds, feet etc.

  

 

  

4.1. 

Fit contact adapter as above. 

4.2. 

Keeping the 

on

 button pressed, now place the contact wheel on the moving surface and read 

the linear rate, ensure wheel is perpendicular to the moving surface. 

4.3. 

Releasing the on button will then hold the last reading in the display for 1 min. 

4.4. 

The instrument retains selected measurement mode for further linear measurements after 
switch off until programmed to a different mode. 

5.  Autorange selection - speed modes only 

5.1. 

While taking a measurement using either up or down measure button, the user can toggle 
between auto and non-auto mode by pressing the programme button, in the auto mode, the 

icon will illuminate. 

6. Average speed monitoring mode -  av icon.

   

 

 

 

6.1. 

Average speed mode

 

- this mode provides a rolling average of the last 8 measured values. 

7. Operation of Maximum & Minimum modes Speed Capture functions - mx, mn. 

 

7.1. 

Having selected the required mode, i.e. Maximum or Minimum, (mx). 

7.2. 

You are now ready to Ca

pture a reading “On Demand” but continuing to operate normally. 

7.3. 

When  a  capture  test  is  ready  to  commence,  while  the  Measure  button  is  held  On,  press 
Programme button   once,  at  this  point  the  instrument  will  switch  into  high  speed  Timebase 
mode,    (0.1  Seconds)  and  will  Capture  the  highest  or  lowest  reading  after  pressing  the 
Programme button. 

 

Releasing the On button will hold the reading and cancel the Capture mode until another 

Capture   measurement is required, when 7.3. should be repeated. 

8.  Count measurement mode - cnt 

 

 

 

 

 

8.1.  

Select mode as described in Measurement mode selection section 

8.2.  

For rev Counting optically, point the light beam at the target and the instrument will measure 
all revs. (pulses) until button is released, the display will hold Count for 1 minute. 

8.3.  

By contact method, fit contact adaptor, press speed cone onto the end of the shaft, the 
Instrument will count revs.

  

9.  Linear Length Totalisation measurements - mt, ft, yd -  

 

 

9.1. 

Select any linear unit of measurement, press contact wheel onto moving surface and 
commence Count by pressing & holding Measure button On, Count will increment until button 
is released. 

9.2. 

The displayed value can be scanned through the equivalent values of Metres, Feet, Yards, 
Count by pressing the Programming button, the instrument automatically calculates the 
appropriate reading. 

Note. Measurement Units will be stored in originally programmed parameter e.g. metres

10.  Time interval measurement - int    

 

 

 

 

10.1. 

Select int mode through Measurement selection mode  

10.2. 

This mode allows measurement of Time between pulses from optical system (or contactor). 

10.3. 

Optically the instrument will measure the time in seconds between pulses, useful for cycle 
timing of reciprocating machinery. 

10.4. 

(Time in seconds per revolution), which equals reciprocal speed. 

10.5. 

Very slow speeds can be measured in this mode below 3 rpm. 

 

11. Display orientation - Inverting function - All modes 

11.1. 

The instrument can be used through 180 deg. rotation (e.g. with the light beam pointing 
downward into a machine), the display inverts so that normal reading can take place.  

11.2. 

The UP Button selects normal mode for optical and contact measurements. 

11.3. 

The Down button selects the display inversion mode and the whole display including relevant 
icons will reverse through 180 degrees allowing access to difficult applications.  

12. Pulse output facility 

12.1 

The A2103/LSR/001 can be connected to external measurement equipment with a TTL 
compatible input. 

12.2 

Connect the lead supplied to the tachometer output socket. The Tachometer will turn on 
automatically and remain powered as long as the lead is attached. 

12.3 

The tachometer will give a 5V TTL compatible signal when the laser is on target. 

 

Reviews: