IE-LGR-UxxxxM-22
65
Error
Description and debugging
Error 20
The source value of the calculated quantities is not available.
Check in the SW whether measurement values are available
(temperature and relative humidity) for this calculated quantity
in question (e.g. the dew-point temperature).
Error 21
Calculation failure, incorrect device calibration.
Have the device repaired.
Error 22
Counter overflow in the case that in the device configuration it is
required to report this as a failure.
Reset the counter by means of the SW, or as the case may be change
the setting for the overflow limit.
Error 23
Thermocouple cold junction temperature measurement error.
Check if the internal temperature measurement channel is turned on for
measurement.
Error 50,
Error 52,
Error 55
Device configuration failure, corrupted configuration.
Read the configuration into the SW, check for the correct setting of all
items, save the configuration in the device again.
Error 51
The Digi/E series probe contains invalid calibration constants.
Get the probe repaired or contact the producer.
Error 56
The value is not defined, the measurement on the given channel is
switched off.
Error 57
The value has not been measured yet, the first measurement is waiting
to be completed.
Error 128
Error 255
Error massages generated during communication with the device.
Repeat the communication, contact the producer.
Annex 2: Connection of the Pt1000/E series probe
connector
Summary of Contents for U0110M
Page 5: ...IE LGR UxxxxM 22 5 Datalogger UxxxxM U0141M model...
Page 60: ...60 IE LGR UxxxxM 22 Dimensions U0110M U8410M U0121M...
Page 61: ...IE LGR UxxxxM 22 61 U0141M U3120M U4440M...
Page 62: ...62 IE LGR UxxxxM 22 U3121M U3631M...
Page 63: ...IE LGR UxxxxM 22 63 U0141TM U0843M U5841M U6841M U7841M...