background image

 

COMET SYSTEM, s.r.o.            Bezrucova  2901      

 

  Specifications are subject to change without notice 

756 61 Roznov pod Radhostem, Czech Republic 

 

IE-LGR-N-Ux8xxM-03 

Dataloggers with GSM modem – UxxxxM family 

QUICK START MANUAL

 

U0843M • U6841M • U7844M 

PRODUCT DESCRIPTION 

 

The dataloggers UxxxxM equipped with a two-part self-locking terminal block are designed for measuring and recording physical and 

electric quantities with an adjustable logging interval (1 sec. to 24 hours). The measured values (instantaneous or the average, minimum 
and maximum values detected during the recording interval) are stored in internal non-volatile memory. The data logging mode can be 

cyclic (when the data memory is completely full, the oldest data are overwrite by the new ones), or non-cyclic (the recording will stop once 
the memory is full). For each measured value it is possible to set two alarm limits. The alarms are signalled by the symbols on the LCD 
display, by flashing the LED, by acoustic or by sending a warning SMS message. The data recording can be performed continuously or 

only when an alarm occurs.   

 

GSM modem is a part of each datalogger. Modem is used to send SMS messages to up to four selected recipients and to send the 
measured values using JSON messages, e.g. to the COMET Cloud. In addition to alarm warning messages, SMS messages containing 

current measured values and alarm statuses can be sent at regular intervals. 

 

Device setting, data downloading and online monitoring is carried out using the computer with the COMET Vision software installed 
(see 

www.cometsystem.com

). The USB interface is used to communicate with the computer.  

 

The datalogger is powered by an internal LiIon battery. The device includes a charging circuit, which is activated automatically when a 
standard USB charger is connected or after connecting a device to a computer. The charging takes place if the battery condition requires it 
and the internal temperature of the device is in the range 0 and 40 ºC.  

 

Device type

 

Measured values

 

Construction

 

 

 

 

U0843M

 

2

 

x

 

Te + 2

 

x

 

BIN

 

Terminal block for connection of two external Pt1000 probe and two binary inputs

 

U6841M   

3

 

x

 

I + 1

 

x

 

BIN 

Terminal block for connection of three current inputs 0 

– 20 mA and one binary input

 

U7844M   

4

 

x

 

BIN 

Terminal block for connection of four binary inputs (two of them can work as a counter)

 

  

 

   Te…Temperature, I…Current 0-20 mA, BIN…Binary inputs 

 

INSTALLATION AND OPERATION

 

 

Insert the micro-SIM card for the 2G network into the device (see other side of this sheet). Use a card with parameters conforming to 
the expected number of sent SMS messages and amount of received and transmitted data. If the SIM card is protected by PIN code, make 
a note it, and later insert this code into the device configuration. Proceed with care and avoid contact of the external conductive parts with 
the electronics of the device (the datalogger is constantly supplied from the internal battery). An optional accessory is a prepaid IoT SIM 

card (order code LP105) for sending data to the COMET Cloud. Please note that this card does not support the sending of SMS messages. 
The UxxxxMsim kit (datalogger with built-in IoT SIM card) allows the instant connection to the COMET Cloud. 

 

Fasten the datalogger on the wall with two screws or insert it into the wall holder LP100 (optional accessory). The device always install 

vertically (with the antenna facing up) into locations with sufficient GSM signal quality. Insufficient signal level can be in reinforced concrete 
buildings, metal chambers and other shielded areas. The device can be operated as a portable – in this case the device protect from falling 

and try to maintain the proper working position. The device should be located as far as possible from potential interference sources. 

 

Connect input cables 

 

Remove the terminal block from the device, connect the wires (maximum wire cross-section 1.5 mm

) with either the supplied SP013 

tool or a screwdriver corresponding to the size and the terminal block again push to the device 

 

The maximum length of input cables should not exceed 30 m (recommended cable length of the Pt1000 probe is 15 m). 

 

All current and binary inputs are not each other galvanically isolated. 

 

The current inputs (U6841M) are passive. The current loops cannot be powered from datalogger directly, external power source 
needed. 

 

The device with all cables should be located as far as possible from potential interference sources 

 

Set-up the device 

 

Connect the datalogger with attached cables to the computer. Use an USB cable with USB-C connector (max. cable length 3 m). 

 

Run the installed COMET Vision software and select the device you want to set up from the device list  

 

Click on the Configuration button. The device configuration will be downloaded and you can change the setup of individual items. 

 

Finally save the new configuration into the device (Apply changes

 

The devices do not require special maintenance. We recommend verifying the measurement accuracy regularly by calibration. 

 

SAFETY INSTRUCTIONS  

 

-   Read carefully the Safety information for dataloggers with GSM modem before operating the device and observe 

it during use! 

-  Installation, electrical connection and commissioning should only be performed by qualified personnel in accordance 

with applicable regulations and standards 

-   Devices contain electronic components, it needs to liquidate them according to currently valid conditions.  
-  To  complement  the  information  in  this  data  sheet  read  the  manuals  and  other  documentation,  which  are 

available in the Download section for a particular device at 

www.cometsystem.com

 