The    probe  for  measuring  the  moisture  of  compressed  air  should  be  placed  directly  on  the 

pressure  pipelines  to  achieve  higher measurement accuracy and fast response times. But  they are 
cases  where  such  placement  is  not  possible.  The  reason  is  the  high  air  speed,  high  temperature, 
high pollution, small diameter pipes, etc. Such situation can be solved by placing the probe into the 
flow  measuring  chamber.  The  picture  shows  the    basic  layout    of    the    sampling  system  with 
chamber SH- PP. 

 

 
 

1 ….sampling 
2 ….closing valve 
3 ….probe 
4 ….flow chamber SH-PP 
5 ….closing valve 
6 ….outlet tube

 

 

 
 
 
 
 
 
 
 

sampling (1)

 - end of the tube placed in the centre of pressure pipelines (distribution of moisture in 

the  pipe  cross-section  is  not  homogeneous).  To  achieve  fast  response  times  to  shorten  the 
length of the sample tubes to a minimum (few meters).

 

closing valve (2) - 

allows access to the sample system without interrupting the main line

 

closing valve (5) - 

the  sample  flow  is  regulated  by  this  valve.  Measurement accuracy is  typically 

not affected by the sample flow rate, but at low speeds, increasing response time.

  

outlet tube (6) -

 

if the measured sample of air is released into the atmosphere, select the length of 

the  outlet  tube  of  1.5  m  (recommended  for  tube  diameter  6mm).  The  reason  is  to  ensure  the 
accuracy  of  the  sample  in  the  flow  chamber  and  avoid  back  diffusion  of  moisture  from  the 
ambient air.

 

That  basic  structure  of  sampling  system  can  be  supplemented  with  filters,  coolers,  flow 

measurement,  pressure  measurement,  etc.  For  the  accurate  operation  of  sampling  system  is 
important  to  ensure  perfect  tightness  of  all  connections  and 
to  use  corrosion-resistant  materials.  Tube  inclination  is 
chosen  so  as  to  avoid  the  accumulation  of  fluid  in  the 
system. 

 

 

Technical specification – flow chamber SH-PP 

 

Material of flow chamber: 

 

stainless

 

steel 

 

 

 

 

 

(DIN 1.4301) 

Inlet and outlet connection:  

G1/8  

Probe connection: 

 

 

G1/2 

Sample flow rate:  

 

 

0.1 to 3 l/min 

Operating pressure: 

 

 

up to 25 bar 

Weight: 

 

 

 

580 g 

 

Note:  

Screw-coupling not included  

Summary of Contents for COMMETER C3121P

Page 1: ...COMMETER C3121P Digital thermometer hygrometer with external probe for compressed measurement Instruction manual...

Page 2: ...element of the device is an external probe see picture designed for compressed air measurement up to 25 bars The probe has a cable length of 1 to 4 meters with connector Mechanical connection of the...

Page 3: ...ge current consumption 0 15 to 0 7 mA depending on operation mode Protection device IP20 sensors are located in cover with IP40 protection Typical life of Zinc Chloride battery 4 months Typical life o...

Page 4: ...ator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument with a...

Page 5: ...ault mode Minimum and maximum memory is cleared from menu after confirmation selection CLR see below Values in HOLD MIN and MAX memories remain stored even after instrument is switched OFF Functions a...

Page 6: ...ication can be activated if menu AUDI On is selected see setting described above Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum This...

Page 7: ...ions and data in memory HOLD MIN and MAX will be lost For the same reason do not disconnect the battery for longer than 1 minute even if instrument is switched OFF If it happens or if battery is total...

Page 8: ...rupting the main line closing valve 5 the sample flow is regulated by this valve Measurement accuracy is typically not affected by the sample flow rate but at low speeds increasing response time outle...

Reviews: