3. If the fraction point is flashed when the matter is taken away, it means that the inner sensor is

unstable. For weighing ordinary run and for weighing precisely run after the fraction point is
stable.

E.-  SUBSTRACTING  TARE

1. Put  a container on a weighing plate, and the mass of the container will be displayed on the scale.
2. Press the button ‘‘TAR’’ and ‘‘0.0’’ will be displayed and the tare has been subracted.
3. Put the matter in the container, and the mass of the matter will be displayed.

F.- COUNTING  OPERATION OF  THE  SCALE

1

Access to set the average state to count  single matter.
Put the container needed in counting at  the weighting plate. If the container is not needed, don’t
put Button  ‘’COU’’, the scale displays ‘’_COU_’’, then  ‘’10’’ flashing is displayed. Then put 10
matters on the weighing plate to be counted, and then press the button  ‘’AFF’’,  for about a
second, the waiting state ‘’_ _ _ _ _ _ _’’ will be displayed. After approximately 3 seconds, ‘‘10’’
will be displayed. The scale will then be operated for counting after average number setting
operation finishes.

2. If the counting requirement is high ideal and the mass uniform for single matter is not, then

choose larger average unit at best. Press button  ‘’COU’’,  then ‘’10’’ flashing is displayed. Press
button ‘’TAR’’, the change from ‘’10, ..., 50, 100, ...,250’’  will be displayed. Choose a value. The
other operation is same as 1.

3. The mass of single matter counted must be larger than 400mg. If less than 400mg, use several

matters as one matter. After counting, transferring counting may be made.

4. Button ‘’COU’’, the scale will go back to weighing.

G.- OVER-WEIGHT ALARM

If the weight of the object to be weighted exceeds ‘’100.5%’’ times, the display will show ‘’-------’’,
to indicate that the total mass to be weighted exceeds the maximum allowed weight range, in
this case, object must be removed. 

Otherwise, the scale will be damaged

.

C.- CALIBRATING THE SCALE

1. The calibrating operation is good and corrects half hour after the scale is warmed up.
2.  Leaving the weighing plate empty, hold the button ‘‘CAL’’ for approximately 3 seconds until

‘’_ _CAL_ _’’ is displayed. When flashing ‘‘1000’’ is displayed, (or antother value depending on
the model) is displayed, put the calibrated weight of 1000g. (or the weigth corresponding to the
scale model) on the weighing plate, the waiting state ‘‘- - - - - - - -’’ is shown on the displayed for
about 10 seconds, then ‘‘1000.0’’ will be displayed. Take the weight away, the scale will go back
to ‘’0, ‘‘0.0’’ or ‘’0.00’’ and the calibration is finished.

    Repeat above-mentioned calibration indications if it can not weight properly.

D.-  WEIGHING

1. The weighing zero point ‘‘0’’, ‘’0.0’’ or ‘’0.00’’ is displayed after the scale is warmed up to be stable

or calibrated.

2. Put the matter to be weighed on the weighing plate, the mass of the matter will be shown at

once.

- 2 -

Summary of Contents for LABORCOM 110-L

Page 1: ... COD 5830025 LABORCOM 210 L COD 5830026 LABORCOM 600 L COD 5830031 LABORCOM 1100 L COD 5830032 LABORCOM 2100 L COD 5830033 LABORCOM 11000 L COD 5830027 COMECTA S A Crta N II km 585 1 08630 Abrera Barcelona Telf 93 7700877 Fax 93 7702362 ...

Page 2: ...e 5ºC 35 ºC Temp Fluctuation Range 5ºC h Humidity Range 50 85 B RUNNING THE SCALE 1 Connect the scale to the power supply of 220V 2 Turn on the power switch The scale will display 8 8 8 8 8 8 then the weighting capacity and then the stabilization signal After this the weighting zero point 0 0 0 or 0 00 will be displayed Now the scale is ready to operate If Error_1 appears turn off the power wait 3...

Page 3: ...0 100 250 will be displayed Choose a value The other operation is same as 1 3 The mass of single matter counted must be larger than 400mg If less than 400mg use several matters as one matter After counting transferring counting may be made 4 Button COU the scale will go back to weighing G OVER WEIGHT ALARM If the weight of the object to be weighted exceeds 100 5 times the display will show to indi...

Page 4: ...guaranteed for one year from the date of adquisiton We recommend you keep the invoice as proof of date of purchase This covers possible defective material of faulty assembly The guarantee does not cover damaged caused by misuse of the apparatus 3 Error Description Treating method E1 CAL value large Calibrate again E2 CAL value small Calibrate again E3 Single matter is less See point F 3 than 400 m...

Reviews: