COMAC CAL FLOW 32 Installation And Technical Conditions Download Page 16

FLOW 32 manual 

 

 

 

   Page 16 (total 18) 

 

  COMAC CAL s.r.o. 

Display 

 

Basic display view  

•  Current flow rate Q [m

3

/h] 

•  Volumetric counter V [m

3

   Both values are shown in 3 decimal places. 
 
 

Display setting view 

 

DO1 digital output: ±Imp / +Imp (bidirectional/unidirectional) 

 

D02 digital  output:  Pos/Neg /  +FS /  Pha /  +Imp /  -Imp  (direction of flow/monitor of 
flow  in  positive  direction/phase-shifted  DO1  /impulses  in  positive  direction/impulses 
in negative direction where the impulse constant is always equal as in DO1) 

 

Flow mFS [l/h] (at DO2 output): switching point in positive direction of flow / 
hysteresis in per cents 

 

AO  analogue  output:  ±Loop/+Loop  (Current  loop  active  during  the  flow  in  both 
directions / in positive direction whereas the setting is specified by DO1 mode) 

 

Limit of current output [l/h]: for 4mA / for 20mA / offset for 4mA / offset for 20mA 

 

Starting flow rate measurement: ±Q [l/h] 

 

Flow direction: Inlet/Outlet (in the direction of the arrow on the meter’s name-plate 

 

Flow  rate  simulation:  +Q  [l/h]  (for  verification  of  the  flow  meter  functionality  and 
connection  to  a  higher-level  system,  after  3  minutes,  the  flow  meter  goes  back 
automatically to measurement mode) 

 

Factory reset: (restoring the meter to factory default state) 

 
Corresponding set point values are displayed according to output pre-selections. Possible 
output combinations correspond to the previous  
Possible output configurations Table. 
 

           Display warning or error views 

•  Excitation circuit error  
•  Impulse output flooding  
•  Unstable measuring signal  
•  Signal out of measuring window 

 
 

Warning  or  error  messages  displayed  are  flashing  in  regular  intervals  and  a  LED  indicator 
according to message type is also lit (Error — red LED, warning — amber LED) and the D03 
digital status output is activated at the same time. 
 
 
If you wish to change parameters, you need to initiate the setup mode within 3 minutes after 
powering  of  the  meter  (the  command  to  modify  setting  is  sent  via  the  communication 
interface, or E-button is pressed and held for approx. 4 seconds). After this period is over, it is 
only possible to view the current settings, the modification of parameters is blocked). Button 
↑ changes parameters and button E confirm values.  

 

 

Summary of Contents for FLOW 32

Page 1: ...FLOW 32 Ver 4 0 Installation and technical conditions Date of issue 18 06 2019...

Page 2: ...line 6 Installation check 8 Wiring system 8 Meter wiring 8 Impulse output OUT1 OUT2 9 Status output OUT3 10 Current output 10 Feasible output configurations 10 Wiring check 11 Putting into operation 1...

Page 3: ...tion design The electronic evaluation unit is incorporated right in the flow meter section in order to form its integral part the respective flow rate sensor grounding electrodes are formed by threade...

Page 4: ...affecting the steady flow must be present in the straight pipeline sections They must be located in the piping after the flow sensor or at the farthest distance before it Sources of disturbances may s...

Page 5: ...take pipe Horizontally laid pipeline Long pipeline Free inlet or outlet Pumps The flow of liquid flow in the flow sensor should be steady and free of whirling For this reason straight sections of pipe...

Page 6: ...the user himself Actual installation in pipeline In case of any intervention in the measuring circuit it is necessary to consider the piping system as being filled up with medium regardless of the sta...

Page 7: ...earth point of the meter with the electrical potential of the earthd so that proper earthing is reached and the meter can operate reliably Electrodes The purity of the electrodes may have an influence...

Page 8: ...electrical equipment When the following operations are performed unprofessionally the claim on warranty for the resulting errors becomes extinct Prior to any manipulation with the meter switch off th...

Page 9: ...of volumetric impulses and for metrological verification of the flow meter Conversion constant and pulse width are set by the manufacturer during production OUT2 can be configured by the manufacturer...

Page 10: ...e current value is directly proportional to that of the flow in the sample The flow limits of the current output are arbitrarily adjustable by the manufacturer during production float data type by mea...

Page 11: ...he supply voltage on the control PCB and stabilization of parameters of the meter takes place afterwards approx 5 sec After that period of time the meter starts measuring Meter status It is continuous...

Page 12: ...ated on the flow meter Safety rules for operator Any interventions in the inductive flow sensor and evaluation unit itself are illegal on the part of operator and they may lead to direct scalding by m...

Page 13: ...DM and Silicone seals PVDF Measuring range Qmin Qmax unidirectional bidirectional for 0 2 12 m s 1 60 Accuracy 1 for 1 10 m s 2 for 0 2 1 m s Repeatability up to 0 4 for 0 1 10 m sec Min medium conduc...

Page 14: ...ow ranges for individual DN sizes Diameter nominal mm Qmin m3 h Qmax m3 h DN 4 0 02 0 5 DN 6 0 03 1 DN 8 0 04 2 DN 10 0 06 3 DN 15 0 2 7 DN 20 0 25 10 Factory settings Impulse constants and current lo...

Page 15: ...iameter nominal mm Threaded connection L1 L2 L3 W H1 H2 H3 K Compact flow meter weight kg 4 161 97 16 5 49 80 70 32 17 1 2 6 161 97 16 5 49 80 70 32 17 1 2 8 161 97 16 5 49 80 70 32 17 1 2 10 161 97 1...

Page 16: ...rification of the flow meter functionality and connection to a higher level system after 3 minutes the flow meter goes back automatically to measurement mode Factory reset restoring the meter to facto...

Page 17: ...extinct Before starting any work on the meter power supply must always be turned off Failures and their symptoms during measurement Unstable indication and outputs may occur upon high proportion of so...

Page 18: ...is used in agreement with this manual the occurrence of faults is very rare Should they ever occur contact our service department If you return the meter to the manufacturing plant adhere to the cond...

Reviews: