3.2  Incoming Inspection 

To  avoid  possibility  of  electrical  shock,  do  not  apply  power  to  the  LISN  if  there  is 
any evidence of shipping damage.  If shipping damage to the product or any of 
the  accessories  is  suspected,  or  if  the  package  contents  are  not  complete, 
contact Com-Power or your Com-Power distributor. 

Please  check  the  contents  of  the  shipment  against  the  package  inventory  in 
section 3.3 to ensure that you have received all applicable items. 

3.3    Package Inventory 

Equipment,  accessories,  and  documents  supplied  with  the  model  LI-1100C  LISN 
are as follows: 

 

Calibration data and Certificate traceable to NIST. 

 

Superior Electric shrouded plugs for EUT and power connection. 

 

Optional items 

 

ISO-17025 calibration data and certificate.  

 

External transient limiter model LIT-153. 

 

 3.4    General Safety Instructions 

The  following  safety  instructions  have  been  included  in  compliance  with  safety 
standard regulations. Please read them carefully.  

 

READ  AND  RETAIN  INSTRUCTIONS

  -  Read  all  safety  and  operating 

instructions  before  operating  the  instrument.    Retain  all  instructions  for 
future reference. 

 

 

HEED  WARNINGS 

-  Adhere  to  all  warnings  on  the  instrument  and 

operating instructions. 

 

 

FOLLOW INSTRUCTIONS 

- Follow all operating and use instructions. 

 

 

WATER AND MOISTURE 

- Do not use the instrument near water. 

 

 

VENTILATION 

-  The  instrument  should  be  used/installed  only  in  locations 

where the flow of air through the ventilation openings is not impeded. 

 

 

MOUNTING  –

  The  instrument  can  be  used  in  Horizontal  or  vertical 

orientation  as  long  as  the  ventilation  holes  are  not  obstructed  and  the 
protective grounding is not defeated.  

 

 

HEAT 

- The instrument should be situated away from heat sources such as 

heat registers or other instruments which produce heat. 

 

Distributed by: Reliant EMC LLC,  +1 408 916

5750,  [email protected], www.reliantemc.com

Summary of Contents for LI-1100C

Page 1: ...LI 1100C LINE IMPEDANCE STABILIZATION NETWORK REV022717 INSTRUCTION MANUAL For LINE IMPEDANCE STABILIZATION NETWORK Model LI 1100C 150 kHz to 30 MHz Distributed by Reliant EMC LLC 1 408 916 5750 info reliantemc com www reliantemc com ...

Page 2: ... 11 6 0 LISN Theory 14 7 0 Typical Performance Data 15 8 0 Warranty 17 8 1 Maintenance 17 List of Figures Figure 1 Single Phase connection with one set of LISN 10 Figure 2 Three Phase connections with two sets of LISN 11 Figure 3 Rear Panel 11 Figure 4 Front Panel 11 Figure 5 Example of equivalent LISN circuit 14 Figure 6 Typical Impedance compared to ANSI C63 5 CISPR 16 1 2 requirements 15 Figure...

Page 3: ...fety precautions operational instructions and warranty information and guidelines and instructions for its proper usage Information contained in this manual is the property of Com Power Corporation It is issued with the understanding that the material may not be reproduced or copied without the express written permission of Com Power Distributed by Reliant EMC LLC 1 408 916 5750 info reliantemc co...

Page 4: ...Immunity Test Systems Impedance Stabilization Networks ISN Line Impedance Stabilization Networks LISN Antenna Masts Near Field Probe Sets Preamplifiers Power Amplifiers Spectrum Analyzers Product Safety Test Equipment Transient Limiters Turntables Antenna Tripods Telecom Test Systems 2 0 PRODUCTS AVAILABLE FROM COM POWER www com power com Distributed by Reliant EMC LLC 1 408 916 5750 info reliante...

Page 5: ...e user that un insulated dangerous voltages are present within the product enclosure and on output connectors These voltages may be of sufficient magnitude to constitute a risk of electric shock to persons The Ground symbol inside a circle indicates terminal which is intended for connection to an external conductor for protection against electric shock in case of a fault or the terminal of a prote...

Page 6: ...rtificate External transient limiter model LIT 153 3 4 General Safety Instructions The following safety instructions have been included in compliance with safety standard regulations Please read them carefully READ AND RETAIN INSTRUCTIONS Read all safety and operating instructions before operating the instrument Retain all instructions for future reference HEED WARNINGS Adhere to all warnings on t...

Page 7: ...nts for electrical equipment for measurement control and laboratory use In addition the basic insulation required for a category I protection device cannot be assured Therefore it is imperative to provide additional measures safeguarding against direct or indirect contact by user The operator is responsible for ensuring that protection is maintained during work with the line impedance stabilizatio...

Page 8: ... or ladder ENVIRONMENTAL CONDITIONS This equipment is designed for indoor use Ambient temperature range during operation should be between 5 C to 40 C STORAGE AND PACKAGING The device should only be stored at a temperature between 25 and 70 C During extended periods of storage protect the device from dust accumulation The original packaging should be used if the device is transported or shipped ag...

Page 9: ...VAC 50 60 Hz Line to Ground 620 VDC Inductor type 50 µH EUT Power connector Shrouded Superior Electric Banana Connector RF measurement connector 50Ω N type female Cooling Forced Air Power Adapter for fan Output 6 VDC 0 5 Amp Input 120 VAC 60 Hz 230 VAC 50 Hz Fuse 1 Amp 250 V Adapter output connector Diameter 2 1 x 5 5 mm Center Positive Mechanical Dimensions of Each LISN L x W x H 21 x 10 x 10 inc...

Page 10: ...nd couples the disturbance voltages to the coaxial measurement port for connection to the measuring instrument like EMI Receiver or Spectrum Analyzer 5 2 Product Setup Before initiating use of LISN ensure that PE conductor is connected to ground The ventilation openings are unobstructed No signal or operating voltages above the permissible limits given in specification The LISN s power input port ...

Page 11: ...a Superior Electric Shrouded pin receptacle for connection to an external DC or AC input power source The LISN is shipped with two matching shrouded socket plug for connection to the input power source cables 5 3 3 5 3 2 5 3 4 Figure 4 Rear Panel Figure 3 Front Panel Figure 2 Three Phase connections with two sets of LISN 5 3 1 5 3 2 5 3 5 5 3 6 5 3 7 Distributed by Reliant EMC LLC 1 408 916 5750 i...

Page 12: ...100C has a Superior Electric Shrouded socket receptacle to power the equipment under test The LISN is shipped with the matching shrouded pin plug for making the connections to EUT power cable 5 3 4 EMI measurement Port The LI 1100C provides a 50 Ω N Type female connector to connecting the Spectrum analyzer or EMI receiver for making conducted noise measurements The shielding connection of the N co...

Page 13: ...n ON before powering the EUT via LISN CAUTION The LI 1100 LISNs uses forced air system to keep the air core inductors cool during the test Always use the supplied DC adapter to power the internal fan during the test Do not obstruct the airflow vents Use the LISNs in a ventilated area Failure to observe these precautions may damage the LISN Overheated LISN may also provide incorrect measurement dat...

Page 14: ...hrough the LISN with minimal attenuation This filter is comprised of a single stage low pass LC filter The inductors L used in the LI 1100C are air core type to eliminate the possibility of saturation and to provide stability The LISN also provides a low impedance path for the RF noise from the EUT to the measuring equipment thereby facilitating the measurement of the EUT RF noise The insertion lo...

Page 15: ...ents Figure 7 Typical Phase compared to CISPR 16 1 2 requirements 20 30 40 50 60 70 0 1 1 10 100 Impedance ohms Frequency MHz LISN Impedance CISPR 16 1 2 Ed 1 2 Tbl 4 50uH Impedance Limits 20 ANSI C63 4 2009 Tbl B 2 50uH Impedance Limits 20 20 10 0 10 20 30 40 50 60 70 0 1 1 10 100 Phase Angle degrees Frequency MHz LISN Phase CISPR 16 1 2 Ed 1 2 Tbl 4 50uH Phase Limits 11 5 Distributed by Reliant ...

Page 16: ...solation compared to CISPR 16 1 2 requirements Figure 9 Typical Insertion Loss 20 30 40 50 60 70 80 90 100 0 1 1 10 100 Isolation dB Frequency MHz LISN Isolation CISPR 16 1 2 Edition 1 2 Isolation Limit Minimum 0 0 2 0 4 0 6 0 8 1 0 1 1 10 100 Insertion Loss dB Frequency MHz LISN Insertion Loss Distributed by Reliant EMC LLC 1 408 916 5750 info reliantemc com www reliantemc com ...

Page 17: ...ectly or indirectly from the use or handling of this product or any inability to use this product separately or in combination with any other equipment When requesting warranty services it is recommended that the original packaging material be used for shipping Damage due to improper packaging will void warranty In the case of repair or complaint Please visit our website www com power com and fill...

Reviews: