background image

3.4. 

After scrubbing, rinse thoroughly with water to 
remove all traces of the detergent.

3.5. 

Coat the entire inside surface with STAT-DRI PLUS 
drying agent. This will induce water/condensate to 
form in an even coat on the inside surface, without 
beading. (The water in contact with the hot cassette 
surfaces also evaporates much more efficiently.)

4.1.

 Ensure mains power lead is attached and plugged in 

and power ON the unit.

4.2.

 

It is recommended that when the first cycle is run, 
and where appropriate to EN13060 compliant cycle 
units, the Statim Helix PCD test device is placed in 
the cassette to check the cycle is operating correctly 
before running the unit with an instrument load.

 

4.3.

 Place the cassette (with PCD) into the unit.

4.4.

 Run a 

“Wrapped Hollow”

 cycle and when the 

sterilisation phase is complete, 

allow AT LEAST 30 

minutes of air drying.

4.5.

 Remove cassette, open, and check the integrity of the 

PCD integrator. 
    

PASS

FAIL

4.6.

 If any residue is evident in the cassette, clean the 

cassette as above and run a second cycle (without PCD). 

4.7.

 If the integrator indicates a pass, the unit is now ready to 

run the appropriate cycle WITH load.

In the unlikely event that the unit faults for any reason that you cannot 
resolve by resetting and running another cycle, or the PCD shows a fail, 
please make a note of the fault and contact your local dealer’s service 
department for advice.

STAT

IM

® 

G4 

Cassette Autoclaves  |   Product Advice Sheet

 

4. FIRST CYCLE AFTER DECOMMISSIONING