OFF
ON
ZERO
MAX
Ultrasonic Scaler QuickStart Guide:
1. Press S button to turn ON the Scaler.
2. Tip Working mode:
3. Select scaler function:
4. Water regulation:
5. Foot Pedal Control:
6. Cleaning:
Scaler
Comprehensive cleaning function:
The cleaning function can be started by pressing
C
. Water
or disinfectant cleans the water line and handpiece. The
cleaning cycle stops automatically after
80 seconds
.
With every push of the
S
button, the function toggles from:
Scaling with water
Dry mode: scaling without water
Irrigation mode: just water, no scaling
The unit is in
standby
mode when the power cord is
connected and the green LED is illuminated.
or
or
100 %
70 %
40 %
Depress the foot control pedal in the leftmost position and
adjust the water flow with the
control ring
on the scaler
handpiece until the water is dripping.
The scaler is
activated
when the foot control pedal is
pressed down. The power is regulated from left/ZERO
to right/MAX within the chosen working mode.
In ZERO
position
there are two functions:
a) irrigation with no scaling power: hold the footpedal
down.
b) diagnostic light: press the footpedal once.
1
1
2
2
3
4
3
4
Choose the correct working mode that is marked on the
base of each tip by pressing the corresponding working
mode button on the keypad:
7. Push the S button for 3 seconds to turn the scaler OFF.