On the top level menu
“
setting
” and “
memory
”
are no longer selectable.
Furthermore, it is not possible
to measure a “
Standard
”.
Standards can only be loaded
from the memory.
To disable the “
User mode
”
select “
Measure
” from the
main menu. In the following
window go on
“Exit user mode
” and
confi rm this with Enter. You
will be prompted to enter the
password.
After entering the correct password, the “
User
Mode
” is canceled and the unit will return to the
measurement menu.
If the “
User mode
” and also the function “Use
Working-Standard” are activated, it is only
possible to calibrate with the Working-Standard.
Warnings
6.1.5 Calibration warning
The Rugged 2 has a calibration warning function. This
informs the user that a recalibration is needed.
Temperature control:
The sensitivity of this function
can be adjusted in “
Settings
”
– “
System
” – “
Calibration
warning
”.
The sensitivity can be selected
as an index of 0 (off ) to 9 (most
sensitive setting). Select the
desired level with the arrow keys
and confi rm with “
Center
”.
Time interval:
A time interval warning since the last calibration can also
be set in intervals up to 48h or switched off .
The sensitivity setting should be adapted to the
required precision – The advised setting for normal
use is 3 and 24h.