Operation

Preparation

1.  Place the aluminium insert blocks to be used into the Dri-Block

®

 unit. The blocks should touch each other  

 

and be positioned centrally.

2.  Loosen the locking knob on the carriage. Remove the gas chamber, disconnecting the gas pipe if necessary,  
 

and lay the chamber upside down on a suitable work surface.

3.  Insert the hyperdermic needles in the pattern to suit the test tubes to be placed in the heater.

 

The needle guide holes are labelled with a letter corresponding with the type of insert block used in the heater. 

 

These are listed in the following table. As an example if you use the block for twelve 15mm test tubes (e.g.  

 

block number F3506) you would place a needle in all the holes labelled C.

 

Only Sample Concentrators with the part number FSC496D or a replacement head part number 6106467 will 

 

have the 96 well array.

Tube size & diameter

Hole array

Hole label

5  x  6 deep

30

A

4  x  5 deep

20

B

3  x  4 deep

12

C

8 hole off-set array, e.g. F4490

D

 

Press each needle firmly, pointed end first, through the guide holes and through the sealing pad behind so  

 

that it is held firmly in position. Build up the desired pattern of needles to suit the arrangement of test tubes  

 

to be used. Do not put a needle in any position which will not have a corresponding sample to be concentrated

 

as this will only waste gas. The needles may be removed and replaced many times as the pad is self sealing.

4.   Replace the gas chamber onto the carriage and tighten the locking knob. Re-connect the gas pipe if necessary.

5

Summary of Contents for Techne FSC400D

Page 1: ...Sample Concentrator FSC400D FSC496D Instruction Manual TECH0002 Version 1 0...

Page 2: ...essing of Samples 8 Maintenance 8 Cleaning and Care 8 Carriage Friction 8 Replacing Sealing Pad 9 In Case of Contamination 9 Service and Repair 10 Warranty 10 Replacement Parts and Accessories 10 Diag...

Page 3: ...into the test tubes The samples are heated from below by the heater and the flow of gas directed over the surface of the samples displaces the evaporated solvent from above the liquid s surface This s...

Page 4: ...e carriage Tighten the locking knob to secure the gas chamber in position 3 Connect the intake nozzle at the rear of the gas chamber to a suitable supply of gas via a pressure reducing valve The gas p...

Page 5: ...F3506 you would place a needle in all the holes labelled C Only Sample Concentrators with the part number FSC496D or a replacement head part number 6106467 will have the 96 well array Tube size diamet...

Page 6: ...FSC400D matrix plate FSC496D matrix plate Inserting the Needles 6...

Page 7: ...the product and could cause a hazard Preventative maintenance should include keeping the product clean by protecting it from spillage contamination or corrosive environments If in doubt please confirm...

Page 8: ...ntamination Certificate Servicing and Repair This product range does not require any routine servicing instructions for cleaning and decontamination are included In the event of product failure it is...

Page 9: ...F7210 Pack of 100 x 127mm long needles FSC4NCS Pack of PTFE coated needles 100 x 76mm long FSC4NCL Pack of PTFE coated needles 100 x 127mm long 6101608 Spigot 6101604 Sealing pad 6101605 Support strip...

Page 10: ...e diameter 6 35 mm Height 530 mm Width 240 mm Depth 295 mm Weight 3 5 Kg Customer and Technical Support For help and support contact Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kin...

Page 11: ...acid Ethanoic acid 118 44 Ethanol Alcohol 79 175 Ethyl acetate Ethyl ester 77 40 Heptane Dipropylmethane 98 120 Hexane n Hexane 69 335 Propan 2 ol Isopropanol 82 137 3 Methyl 1 butanol Isoamyl alcohol...

Page 12: ...g the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 330 088 0560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specific...

Reviews: