background image

Product description

 

20 

 

Mains connection 

 

 

100-230 V 

50/60 Hz 

 

 

Nominal current consumption 

5 (100 V) 

3 (230 V) 

 

 

Total current consumption 

16 

 

 

Mains fuse, resettable 

10 

 

 

 

*00 

 at 100 V/ 

Technical data 

900F 

Performance values 

Lowest temperature 

°C 

-38/-40* 

Refrigerants 

 

R449A 

Dimensions 

Dimensions (W x D x H) 

cm 

39 x 62 x 53 

Usable bath opening (W x D) 

cm 

26 x 35 

Bath depth 

cm 

20 

Volumes min. … max. 

21.0 … 30.0 

Weight 

kg 

47 

Mains connection 

 

 

115 V 
60 Hz 

200-230 V 
50/60 Hz 

 

Nominal current consumption 

8   

5 (200 V) 

5 (230 V) 

 

Total current consumption 

16 

16 

 

 

Switzerland 

 

10 

 

 

Great Britain 

 

13 

 

Mains fuse, resettable 

14 

10 

 

* depending on the circulator used 

 

* at 100 V/ 

 

 

Summary of Contents for Julabo 1000F

Page 1: ...Refrigerated bath basic units Original operating manual 1 950 0170 us V08 10 2022...

Page 2: ...nt of this operating manual is protected by copyright Information including texts images and other contents may not be reproduced distributed transmitted stored or otherwise used in any form without p...

Page 3: ...binations with circulators 11 5 3 Operating and functional elements 12 5 3 1 External inverter 13 5 4 Technical data 14 5 4 1 Material of parts that come into contact with the medium 22 5 4 2 Refriger...

Page 4: ...ent choice JULABO would like to thank you for the trust you have placed in our company and products This operating manual will help you become acquainted with the use of our units Read the operating m...

Page 5: ...n the operating manual of the associated circulator 2 2 Original JULABO spare parts Hassle free continuous operation and safety also depend on the quality of the spare parts used Only original JULABO...

Page 6: ...a danger with a low level of risk which if it not prevented may result in minor to moderate injuries WARNING This signal word designates a danger with a medium level of risk which if it not prevented...

Page 7: ...f the unit so that the operator can operate the unit safely and avoid misuse This is a laboratory device intended for temperature control applications of liquid media It can only be operated in conjun...

Page 8: ...nd fluids Wear suitable protective gloves Electric shock from electrical system Touching damaged live parts can cause severe electric shocks and lead to injury or even death Have damaged insulation an...

Page 9: ...fety shoes Protective clothing Breathing protection Hearing protection Face and eye protection Specify and provide personal protective equipment for the respective application Use only personal protec...

Page 10: ...unctions are not affected by the control circuit If a safety function trips all actuators are permanently switched off The operator is warned by an optical and acoustic alarm on the circulator High pr...

Page 11: ...cisely cool and heat samples over a wide temperature range The device combination is suitable for external temperature applications The cooling machine is controlled by the connected circulator 5 2 Po...

Page 12: ...their position on the unit Front without cover and rear e g 310F 1 Recessed grip 2 Bath tank drain opening 3 Drain valve 4 Drain hose for 310F and 450F 5 Condenser 6 Recessed grip 7 CAN plug for conn...

Page 13: ...0 V voltage variant of 310F and 450F the inverter is mounted on the rear side Rear side of 310F 450F 100 V voltage variant 1 Inverter 2 CAN plug for connection with a circulator 3 Mains input socket 4...

Page 14: ...s A Use in an industrial electromagnetic environment In accordance with IEC 61010 1 the device is designed for safe operation under the following ambient conditions Indoor use Altitude up to 2000 m ab...

Page 15: ...ght kg 23 Mains connection 100 V 50 60 Hz 115 V 60 Hz 230 V 50 60 Hz Nominal current consumption A 4 4 2 Total current consumption A 15 12 16 Switzerland A 10 Great Britain A 13 Mains fuse resettable...

Page 16: ...Performance values Lowest temperature C 25 Refrigerants R134a Dimensions Dimensions W x D x H cm 24 x 42 x 66 Usable bath opening W x D cm 13 x 15 Bath depth cm 15 Volumes min max l 3 0 4 0 Weight kg...

Page 17: ...3 Mains connection 100 V 50 60 Hz 115 V 60 Hz 200 230 V 50 60 Hz Nominal current consumption A 4 4 2 200 V 3 230 V Total current consumption A 15 12 16 Switzerland A 10 Great Britain A 13 Mains fuse r...

Page 18: ...0F Performance values Lowest temperature C 35 Refrigerants R449A R452A Dimensions Dimensions W x D x H cm 33 x 47 x 47 Usable bath opening W x D cm 22 x 15 Bath depth cm 15 Volumes min max l 5 0 7 5 W...

Page 19: ...00 V 50 60 Hz 115 V 60 Hz 200 230 V 50 60 Hz Nominal current consumption A 11 7 3 200 V 4 230 V Total current consumption A 15 12 16 Switzerland A 10 Great Britain A 13 Mains fuse resettable A 12 12 1...

Page 20: ...t temperature C 38 40 Refrigerants R449A Dimensions Dimensions W x D x H cm 39 x 62 x 53 Usable bath opening W x D cm 26 x 35 Bath depth cm 20 Volumes min max l 21 0 30 0 Weight kg 47 Mains connection...

Page 21: ...onnection 115 V 60 Hz 200 230 V 50 60 Hz Nominal current consumption A 9 6 200 V 6 230 V Total current consumption A 16 16 Switzerland A 10 Great Britain A 13 Mains fuse resettable A 14 10 depending o...

Page 22: ...bath fluid used Parts that come into contact with the medium Material Bath tank 1 4301 304H Bath tank drain opening 1 4301 304H Bath tank bath cover gasket FKM Viton O ring drain valve FKM Viton 5 4 2...

Page 23: ...evice against tipping in the center of the transport trolley 4 Place loose parts for the device such as cables on the transport trolley The device is then ready for transport and can be safely transpo...

Page 24: ...ements Insufficiently dimensioned power cords must not be used See type plate for mains voltage and current value We recommend only using original JULABO spare parts 7 2 Check safety symbols The follo...

Page 25: ...gns order no 3 383 2630 3 Note order no 3 383 2640 1 Check the safety symbols on the unit for legibility and completeness 2 Replace defective or missing safety symbols Safety symbols can be reordered...

Page 26: ...g Before draining the device let it cool to room temperature Avoid direct contact with hot bath fluid Wear protective gloves The device is tempered to room temperature and switched off 1 Place a suita...

Page 27: ...chnical service for alternative cleaning mediums The device has now been cleaned 7 5 Cleaning condenser From time to time the condenser on the front of the device should be cleaned to maintain full co...

Page 28: ...vice for repair Follow the procedure described to ensure that your device continues to function reliably even after being stored for a long period The device is switched off and disconnected from the...

Page 29: ...nical Service Tel 1 610 231 0250 Option 3 Fax 1 610 231 260 Email Service julabo us Before sending a device to Technical Service the following points must be observed Clean and decontaminate the devic...

Page 30: ...not include damage to the product resulting from accident misuse improper installation or operation unauthorized or improper repair replacement or alteration including but not limited to repairs repl...

Page 31: ...th above Seller shall have no further liability or obligation to Buyer therefor In no event shall Seller s liability under the Initial Warranty exceed the original purchase price of the product which...

Page 32: ...to any other provision of the Warranty Provisions nor as to any prior or subsequent breach of the same provision Freight Seller will arrange and pay for shipping and handling for the return of the un...

Page 33: ...e with national or regional regulations The circulator combination is switched off and disconnected from the mains voltage 1 Allow the device to cool down to room temperature 2 Disconnect all power ca...

Page 34: ......

Page 35: ......

Page 36: ......

Reviews: