background image

5.6 Abbe condenser 

Beneath the object stage an Abbe condenser N.A. 12.5 is mounted. The condenser is factory pre-centered but can be 
adjusted in height by means of a rack and pinion movement and knob. With this, one can focus the light on the specimen 
by which the contrast can be optimized. If needed the following procedure can be used to center the condenser:

1. 

Move the condenser to the highest position 

2. 

Select the 10x objective into the light path and focus the specimen

3. 

Rotate the field diaphragm adjustment ring to put the field diaphragm to the smallest position 

4. 

Rotate the condenser up/down knob, and adjust the image to be clearest 

5. 

Adjust the center adjustment screw and put the image to the center of the field of view 

6. 

 Open the field diaphragm gradually. If the image is in the center all the time and inscribed to the field of view,  
the condenser has been centered correctly

5.7 The field (Köhler) diaphragm (A)

By limiting the diameter of the beam entering the condenser, the 
field diaphragm can prevent stray light and enhance the image 
contrast. When the image is just on the edge of the field of view, the 
objective can show the best performance and obtain the clearest 
image. The diaphragm is factory pre-centered.

5.8 Adjusting the Aperture Diaphragm (B)

• 

The diaphragm is used to adjust the Numerical Aperture of the 
condenser. When the N.A. of the condenser matches the N.A. 
of the objective, the highest possible resolution is obtained

• 

When contrast is low, rotate the diaphragm adjustment ring 
to 70% to 80% of the N.A. of objective. This will improve the 
contrast of the image. The diaphragm is factory pre-centered.

5.9 Use of the S100x oil-immersion objective

The MSU-600 Series is equipped with an S100x N.A. 1.25 oil 
immersion objective. Please follow these instructions for use:

1. 

Remove the dust protection from the revolving nosepiece  
to mount the S100x objective

2. 

Focus the image with the S40x objective

3. 

Turn the revolving nosepiece so the S100x objective almost reaches the click-stop

4. 

Put a small drop of immersion oil on the center of the slide

5. 

Now turn the S100x objective so that you feel the click stop

6. 

The front lens is in contact with the immersion oil

7. 

Look through the eyepiece and focus the image with the fine adjustment knobs

8. 

The distance between the lens of the objective and the slide is very small!

9. 

 In case there are small bubbles visible turn the S100x objective a couple of times left/right so that the front of  
the objective moves in the oil and the bubbles will disappear

10. 

 After using the S100x objective turn the table with the fine adjustment knobs downwards until the front lens 
doesn’t touch the oil any longer

11. 

 Always clean the front lens of the S100x objective with a lens paper moistened with a drop of isopropanol

12. 

Clean the slide after use as well

5.10 Safety device

To prevent damage to the objective lens or the slide, all types are equipped with a pre-fixed safety device. It is 
recommended to use slides of 1.0 – 1.2 mm thickness in combination with cover glasses of 0.13 mm or 0.17 mm thickness.

12

Summary of Contents for 78904-16

Page 1: ...Installation and Operating Instructions MSU 600 Series Compound Microscopes Models 78904 16 18 21 24 27 30 33...

Page 2: ...____________________________________________________________________ 13 5 12 Auto Off Function ______________________________________________________________________ 13 6 0 Phase contrast model 78904...

Page 3: ...enclosed with a cover glass never comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated...

Page 4: ...or non alcohol based disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer a RMA return authorization form and a decontamination...

Page 5: ...se undue force when turning the knobs Make sure that the microscope system can dissipate its heat fire hazard Keep the microscope approximately 15 cm free from walls and obstructions Never turn the mi...

Page 6: ...04 16 Binocular 4x 10x S40x S100x semi plan 48 to 76 mm Rackless integrated X Y mechanical stage 3 W NeoLED 100 240 VAC 50 60 Hz 24 1 lb 10 9 kg 78904 18 Trinocular 4x 10x S40x S100x semi plan 48 to 7...

Page 7: ...e J Nosepiece for 5 objectives C Microscope head K Objectives D Height adjustment condenser L Stage with X Y mechanical stage E Slide protection handle M Condenser with iris diaphragm F Coaxial coarse...

Page 8: ...sure the power switch is turned off the plug removed before replacing the led unit or fuse 4 1 Assembling steps The steps mentioned below are often not necessary but described for your convenience no...

Page 9: ...l Each eyepiece has a rubber eyeshade This prevents damage to the lens and stray light The eyeshade can simply be slipped over the eyepiece Connecting the power cord The MSU 600 series supports a wide...

Page 10: ...of the specimen holder backwards Release the arm slowly clamping the slide with the cover glass facing up Rotating the X and Y axis knob will move the specimen to the center for alignment with the ce...

Page 11: ...one diopter adjustment Position the diopter marking on the zero point Look into both eyepieces and focus on the specimen Close the right eye and look into the eyepiece with diopter adjustment rotate...

Page 12: ...ion is obtained When contrast is low rotate the diaphragm adjustment ring to 70 to 80 of the N A of objective This will improve the contrast of the image The diaphragm is factory pre centered 5 9 Use...

Page 13: ...image appears in the eyepiece 6 2 Using the Phase contrast slider Keep the phase contrast slider face up text up insert it from left to right into the condenser slider socket as the direction of the...

Page 14: ...w If the ring images can t be observed clearly first try and focus the centering telescope If this does not solve the issue raise or decline the condenser If the bright ring and dark ring images are n...

Page 15: ...blow in order to prevent internal damage to the microscope And in most cases replacing with a correctly rated fuse will resolve the issue However should you encounter a blown fuse frequently please co...

Page 16: ...orm the height adjustment procedure above while watching the image on the computer screen In this case once you have obtained parfocality in the device tighten screw A again Screw B is only used to fi...

Page 17: ...___________________________________________________ ___________________________________________________________________________________________________ ________________________________________________...

Page 18: ...11 0 Supplementary Manuals 11 1 Supplementary Manual for Darkfield Model 78904 27 18...

Page 19: ...ndenser by unwinding the screw marked with the arrow Set up the dark field condenser and tighten the screw 3 Lower the condenser slightly place the slide In this way the immersion oil should not yet t...

Page 20: ...is aperture you must also put immersion oil on the prepared slided 9 Focus the lens On the S100x 1 25 oil immersion objective A and S100x 1 25 IOS super contrast objective B you can adjust the backgro...

Page 21: ...11 2 Supplementary Manual for Metallurgical Model 78904 30 21...

Page 22: ...l Model 78904 30 cont Components A Microscope head B Aperture diaphragm C Field diaphragm D Two slots for color filters E Condenser focusing knob F Brightness adjustment knob G Main body of metallurgi...

Page 23: ...ing the Illumination 1 Adjust the condenser knob 1 E to the position where the image brightness in view is homogenous 2 Adjust the field diaphragm 1 C and aperture diaphragm 2 B until the two diaphrag...

Page 24: ...11 3 Supplementary Manual for Polarization Model 78904 33 24...

Page 25: ...e rotation clamping knob 1 is loosened the stage can be rotated 360 horizontally Using the Bertrand Lens By revolving the Bertrand lens dial 3 the Bertrand lens can be selected At the O position the l...

Page 26: ...de approximately half way to the eyepiece crosshairs c 9 Rotate the stage back to its 0 position 10 The 4x objective is equipped with two adjustable screws inside the revolving nosepiece for centering...

Page 27: ...n be rotated 360 degrees turn it until complete extinction is obtained 5 Place the specimen for orthoscopic observation 6 Insert test plates for further observation test and study Conoscopic observati...

Page 28: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 600_Series_Compound Microscopes_Manual...

Reviews: