Care and Maintenance

The servicing instructions here are intended for limited maintenance. Do not perform  

any servicing other than that contained in the instructions. Return product to factory for  

maintenance not covered in this section. Maintenance should only be undertaken by a  

qualified electrician

1.  Housing and Filter Maintenance

The plastic housing is designed for protecting the internal parts and facilitates handling. Make 

sure no liquids come into contact with the electrical components. Clean the filter with a glass 

cleaner and soft cloth.

2.  Tube and Reflector Maintenance

Refer to the following instruction in the “Tube Replacement” section below and remove tubes 

before cleaning. Glass cleaner may be used for cleaning the reflector and a mild detergent 

may be used for cleaning the tube. Wipe and dry with soft cloth.

3.  Tube Replacement 

(tube type TUV 6W)

When the product fails to operate correctly, check with above operating instructions first.  

If the problem persists and tube replacement is deemed necessary, proceed as follows:
1. Turn the power switch to OFF position and unplug the power cable from electrical outlet.
2.  Lay the UV lamp housing on a flat table.
3.  Remove the 8 screws at the filter’s side of the plastic housing. Separate the top and  

bottom housing parts.

Always unplug the product from its power source before cleaning or servicing.

WARNING

Operating Instructions

1. Connecting Power Supply

 

Refer to Specifications on page 6.

2. Turn ON the Unit

 

For Models with Single wavelength: Press power switch to position “I”.

 

 For Models with Dual Wavelength: Select desired wavelength by pressing the select 

switch at right. Press power switch to position “I”.

 

Note:

 The intensity of the radiation might be weak initially while the lamp tube is cold.

3. Turn Off

 

Press power switch to position “O”.

4. Charging Battery for 12 VDC Models

 

 Each 12 VDC operated model contains a lithium battery pack. The battery needs to be 

charged when the yellow status light is on. Please use the charger which comes with the 

product.

Summary of Contents for 09818-05

Page 1: ...Ultraviolet Hand Lamps Operating Manual coleparmer com ...

Page 2: ...y before operating the product Use the product according to the instructions otherwise the protection provided by the product may be impaired WARNING ATTENTION ULTRAVIOLET RADIATION IS HARMFUL TO THE HUMAN BODY Function These UV Hand Lamps provide UV radiation with wave lengths of 365 nm 254 nm or a combination of 254 nm and 365 nm They are used for detecting nucleic acids stained with fluorescent...

Page 3: ...m The rechargeable unit uses 12 V rechargeable batteries and comes with a charger 100 240 VAC 50 60 Hz When the battery is fully charged the unit can sustain about 3 hours continuous use or approximately 500 intermittent uses The battery will have sufficient power to operate when the indicator light is green The battery will need to be recharged if the indicator light is yellow It will take approx...

Page 4: ...ly check with above operating instructions first If the problem persists and tube replacement is deemed necessary proceed as follows 1 Turn the power switch to OFF position and unplug the power cable from electrical outlet 2 Lay the UV lamp housing on a flat table 3 Remove the 8 screws at the filter s side of the plastic housing Separate the top and bottom housing parts Always unplug the product f...

Page 5: ...following step 6 under Tube Replacement 5 Care and Maintenance for Battery The following care and maintenance is only for models with rechargeable batteries 1 The battery charger comes with the unit It works with AC 100 240 V 50 60 Hz 2 It takes up to 4 hours to recharge the lithium battery completely 3 The battery charger and lithium battery system in these models have a built in self protection ...

Page 6: ...Specifications Lithium Battery Operated Models with DC 12 V 0 3 A Wavelength Number of UV Tubes Plug US Universal UV A UV C Regular Battery Operated Models 09818 12 09818 27 UV C 2 09818 13 09818 28 UV A 2 09818 14 09818 29 UV A UV C 1 1 Models with Automatic Safety Switch 09818 08 09818 24 UV C 2 09818 09 09818 25 UV A 2 09818 11 09818 26 UV A UV C 1 1 Long wave UV A 365 nm Short wave UV C 254 nm...

Page 7: ... to qualify for the warranty repair the customer must submit all warranty information required to validate proof of purchase Visit coleparmer com product warranty registration to submit information Should warranty service be required repair or replacement will be provided by Cole Parmer without charge Freight to ship the unit to Cole Parmer will be the responsibility of the purchaser There are no ...

Page 8: ...IL 60061 Toll free 1 800 323 4340 International 1 847 549 7600 sales coleparmer com coleparmer com In Canada 1 514 355 6100 In China 86 21 51099909 In France 33 0 1486 37800 In Germany 49 0 9377 92030 In India 591 22 61394444 In Italy 39 0 1313 89513 In UK 44 0 1480 272279 ...

Reviews: