In-Sight
®
1700 Series Wafer Reader
Section 7: Native Mode Commands
67
7.2 Put Extended Native Mode Commands
The following commands are used in conjunction with the Put extended Native Mode
command to send information to the In-Sight spreadsheet and wafer reader. The
information sent is dependent on the command and input that is specified.
Syntax
: Put [command]
7.2.1 Put Live
Put Live(<Input>)
Turns live acquisition on/off.
NOTES
■
The wafer reader must be Offline when using this command.
■
The live image will not be displayed in the GUI or PC Host.
Inputs
:
0
to turn off live acquisition mode, and
1
to turn on live acquisition mode.
Output
: Status Code, # of bytes
The status codes are:
STATUS CODE
DESCRIPTION
1
The command was executed successfully.
0 Unrecognized
command.
-1
The input is invalid.
-2
The command could not be executed, or the wafer reader is Online.
Sample Input and Output
NOTE
In this sample, the first command puts the wafer reader into live acquisition mode, and the second
command takes the wafer reader out of live mode.
PUT LIVE 1
1
PUT LIVE 0
1
Summary of Contents for In-Sight 1721
Page 1: ...IN SIGHT 1700 SERIES WAFER READER USER MANUAL...
Page 2: ......
Page 8: ...In Sight 1700 Series Wafer Reader vi...
Page 58: ...Section 4 Managing Job Files In Sight 1700 Series Wafer Reader 48...
Page 130: ...Section 8 Specifications In Sight 1700 Series Wafer Reader 120...
Page 136: ...Updating the Wafer Reader s Firmware In Sight 1700 Series Wafer Reader 126...
Page 140: ...Configuring Microsoft Windows Network Settings In Sight 1700 Series Wafer Reader 130...
Page 154: ...In Sight 1700 Wafer Reader Mount Conversion In Sight 1700 Series Wafer Reader 144...
Page 156: ...In Sight 1700 Series Wafer Reader www cognex com products InSight default asp Rev 16 May 2006...