background image

    OPERATION SUMMARY: 

1. Insert a particulate filter onto the end of the sample probe and then insert a  needle onto the
filter.

2. Press the POWER switch to start unit. No warm up is required.

3. Press the PUMP switch. The pump will come on for approximately 4 to 6 seconds and pull
room air into the sensor through the syringe needle and filter. Note that the meter reading will
decrease slightly when the pump is on. This is normal due to the slight vacuum created in the
sensor when the pump is running. Wait approximately 10-12 seconds for the reading to
stabilize.  Room air should give a reading of 20.7 to 21.1 % oxygen.

4. Pierce the food pack with the syringe needle, preferably through  a foam rubber septum
(supplied). Depress the PUMP switch and wait  for the reading to stabilize (10-12 seconds).
Leave the needle in the package until a stable reading has been obtained.

   NOTE: It is difficult to get accurate readings when piercing the bag directly with the needle 
because air will frequently leak in around the hole caused by the needle. This will result in 
inaccurate high readings. This is prevented by first inserting a  foam seal pad onto the package 
surface and piercing the needle through the foam. 

CAUTION: If using alkaline batteries, do not plug the charger into the unit.  These batteries 
can overheat or crack if an extrnal voltage is applied and can melt the battery compartment. 
The unit is fuse protected to prevent this. If the unit will not operate under battery power 
alone but does operate with the charger plugged in , the fuse is probably blown. 

CALIBRATION SETTING 

Accurate SPAN calibration of the unit can be checked at any time by sampling room air, 
which should give a reading of 20.9% oxygen, plus or minus 0.4%. If the reading is off, it can 
be set by adjusting the SPAN potentiometer, located on the side of the unit, with the adjusting 
tool supplied with the unit. 

ZERO SETTING 

The zero setting of the O2 channel is very stable, and will change little even over a period of 
several months. Although usually not required, The O2 ZERO reading can be checked by 
filling a plastic bag with nitrogen and then drawing out a sample as described previously. 
Flush the bag thoroughly first because it is difficult to expell all the air out of it to get a true 
zero sample. The zero will read 0.0, plus or minus 0.1% O2. If the reading is not in this range 
adjust the O2 ZERO potentiometer as needed. Several turns will be needed to see any 
movement in the reading, as the adjustment is not very sensitive.This adjustment should be 
done carefully because it will affect all readings, and it must be done with N2 in the sensor. 

Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL (804) 435-5522 

www.coffeelabequipment.com

Summary of Contents for 901

Page 1: ...t internal sampling pump one main circuit board battery compartment NiCad rechargeable batteries battery charger module and sampling probe The sample probe is tipped with a particulate filter and luer fit syringe needle with side port holes to prevent plugging Sample is drawn through the probe and tubing and then to the oxygen sensor when the pump is turned on The pump is electronically timed to d...

Page 2: ...the unit These batteries can overheat or crack if an extrnal voltage is applied and can melt the battery compartment The unit is fuse protected to prevent this If the unit will not operate under battery power alone but does operate with the charger plugged in the fuse is probably blown CALIBRATION SETTING Accurate SPAN calibration of the unit can be checked at any time by sampling room air which s...

Page 3: ...pump if there are no leaks With this positive pressure air cannot leak into the system with any leaks which may be present If the reading goes lower it indicates a leak somewhere which should be corrected PUMP TIME The pump run time can be increased or decreased as needed by adjusting the PUMP TIME potentiomenter on the side of the unit Be careful not to set the time too high with samples that hav...

Page 4: ...ool to tighten because this can strip the threads BATTERY OPERATION AND RE CHARGING The Ni Cd batteries in the 901 require several charge discharge cycles before reaching full capacity During the first 40 50 hours of use the battery charge will not be optimum but will improve as the unit is used To properly condition the batteries it is important during this initial period of use to run the unit i...

Page 5: ...Ni Cd batteries designed to accept a quicker charge can also be used as long as they are AA size 1 2 1 4 volt and are of the nickel cadmium type Do not use other rechargeable types such as lithium AA alkaline non rechargeable batteries can be used in some situations but extreme caution must be exercised because plugging the unit into the charger with alkaline batteries installed can cause overheat...

Page 6: ...idle Charger included Battery Life Eight hours before re charging Size 8 x 4 x 1 8 in 200 x 100 x 40 mm Weight 1 lb 450 gms OPERATING SUPPLIES 1 9001 Pkg of 200 foam seal pads 2 9002 Package of 1000 foam seal pads 3 9005 Particulate filter pkg of 5 4 9028 Particulate filter pkg of 25 5 9003 Side Port Needles pkg of 3 6 9029 Side Port Needles pkg of 15 7 9007 Sample probe assembly 8 9011 Battery Ch...

Page 7: ...filter provided which fits onto the end of the sampling probe Avoid direct sampling of pressurized gas cylinders High pressure can cause damage to the sensor To test the content of a cylinder flush the sample into a plastic bag and sample from the bag Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL 804 435 5522 www coffeelabequipment com ...

Reviews: