MODEL 901 OPERATING INSTRUCTIONS 

GENERAL DESCRIPTION 

The Quantek Model 901 is a battery-operated, portable oxygen analyzer used for the 
measurement of residual oxygen in gas-flushed (CAP/MAP) food packages.  The Model 901 
is powered by Ni-Cad  rechargeable batteries, with typically 8 hours of operation before 
recharging is required. 

The Model 901 design has proven itself in hundreds of installations for food applications such 
as meats, fresh produce, pasta, nuts, snackfoods, dairy products and coffee. The sensor is 
unaffected by high levels  of background gases such as nitrogen or carbon dioxide. The Model 
901 can also be used for applications where the oxygen level is very high (50 to 100%), 
without overloading the sensor. It is unaffected by trace hydrocarbons or high levels of CO

2

,

 

which can poison other types of sensors. 

Components which make up the Model 901 include the case, oxygen sensor, LCD readout, 
internal sampling pump, one main circuit board, battery compartment, NiCad rechargeable 
batteries, battery charger module and sampling probe. The  sample probe is tipped with a 
particulate filter and luer fit syringe needle with side-port holes to prevent plugging. Sample is 
drawn through the probe and tubing and then to the oxygen sensor when the pump is turned 
on. The pump is electronically timed to draw in the precise amount of sample required for the 
analysis, and then turns itself off after the pre-set sampling time, which is adjustable from 1 to 
10 seconds.  The oxygen concentration is read on the display.

The operating controls and adjustments include power switch, pump switch, span, zero and 
pump time adjustments. 

PRECAUTIONS 

Follow these guidelines to prevent damage to the unit. 

1. Do not suck liquids into the unit through the needle.
2. Do not plug the unit into the charger if alkaline batteries are installed.
3. Do not immerse in water or spray water on the unit
4. Do not drop the unit

Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL (804) 435-5522 

www.coffeelabequipment.com

Summary of Contents for 901

Page 1: ...t internal sampling pump one main circuit board battery compartment NiCad rechargeable batteries battery charger module and sampling probe The sample probe is tipped with a particulate filter and luer fit syringe needle with side port holes to prevent plugging Sample is drawn through the probe and tubing and then to the oxygen sensor when the pump is turned on The pump is electronically timed to d...

Page 2: ...the unit These batteries can overheat or crack if an extrnal voltage is applied and can melt the battery compartment The unit is fuse protected to prevent this If the unit will not operate under battery power alone but does operate with the charger plugged in the fuse is probably blown CALIBRATION SETTING Accurate SPAN calibration of the unit can be checked at any time by sampling room air which s...

Page 3: ...pump if there are no leaks With this positive pressure air cannot leak into the system with any leaks which may be present If the reading goes lower it indicates a leak somewhere which should be corrected PUMP TIME The pump run time can be increased or decreased as needed by adjusting the PUMP TIME potentiomenter on the side of the unit Be careful not to set the time too high with samples that hav...

Page 4: ...ool to tighten because this can strip the threads BATTERY OPERATION AND RE CHARGING The Ni Cd batteries in the 901 require several charge discharge cycles before reaching full capacity During the first 40 50 hours of use the battery charge will not be optimum but will improve as the unit is used To properly condition the batteries it is important during this initial period of use to run the unit i...

Page 5: ...Ni Cd batteries designed to accept a quicker charge can also be used as long as they are AA size 1 2 1 4 volt and are of the nickel cadmium type Do not use other rechargeable types such as lithium AA alkaline non rechargeable batteries can be used in some situations but extreme caution must be exercised because plugging the unit into the charger with alkaline batteries installed can cause overheat...

Page 6: ...idle Charger included Battery Life Eight hours before re charging Size 8 x 4 x 1 8 in 200 x 100 x 40 mm Weight 1 lb 450 gms OPERATING SUPPLIES 1 9001 Pkg of 200 foam seal pads 2 9002 Package of 1000 foam seal pads 3 9005 Particulate filter pkg of 5 4 9028 Particulate filter pkg of 25 5 9003 Side Port Needles pkg of 3 6 9029 Side Port Needles pkg of 15 7 9007 Sample probe assembly 8 9011 Battery Ch...

Page 7: ...filter provided which fits onto the end of the sampling probe Avoid direct sampling of pressurized gas cylinders High pressure can cause damage to the sensor To test the content of a cylinder flush the sample into a plastic bag and sample from the bag Coffee Laboratory 589 Rappahannock Drive White Stone Va 22578 TEL 804 435 5522 www coffeelabequipment com ...

Reviews: