is likely to cause interference, in which case the user, at his own expense, will be required to take whatever
measures may be appropriate to correct the interference. If your device does cause interference to radio or
television reception, you are encouraged to try to correct the interference by one or more of the following
measures:
The main difference between this document and the last was that all bulleted lists have had the “List
Bullet” style applied. This is different than the “List Paragraph” style that is applied by default. With this
change bulleted lists are copied over properly.
Reorient the receiving antenna
Relocate the device with respect to the receiver
If necessary, you should consult Codonics Technical Support or an experienced radio/television technician for
additional suggestions. You may find the following booklet prepared by the Federal Communications Commission
helpful: How to Identify and Resolve Radio-TV Interference Problems. This booklet is available from the U.S.
Government Printing Office, Washington, D.C. 20402, Stock No. 004-000-00345-4.
This product is in conformity with the protection requirements of EC Council directive 89/336/EEC on the
approximation of the laws of the Member States relating to electromagnetic compatibility. This product satisfies
the Class B limits of EN55011. A declaration of conformity with the requirements of the Directive has been signed
by the Director of Quality Assurance and Regulatory Affairs.
Guidance Regarding Electromagnetic Emissions and Immunity
Suitable Environments:
The Codonics Virtua Medical Disc Publisher is intended for use in professional healthcare facility environments,
including hospitals and medical clinics.
The Codonics Virtua Medical Disc Publisher has not been evaluated for use near HF surgical equipment. If use
near HF surgical equipment is desired, the user is responsible for verifying proper operation of the Virtua. If
Virtua does not perform correctly in this environment, move the Virtua farther from the source of the
electromagnetic disturbance.
The Codonics Virtua Medical Disc Publisher has not been evaluated for use in emergency medical vehicles.
As a support device, the Codonics Virtua Medical Disc Publisher does not provide essential performance.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally
WARNING
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Virtua, its cables, or accessories.
Otherwise, degradation of the performance of this equipment could result.
Electromagnetic Emissions Standards and Test Levels: