![CODEX DNA BioXp 3200 system Installation Manual Download Page 9](http://html.mh-extra.com/html/codex-dna/bioxp-3200-system/bioxp-3200-system_installation-manual_2634123009.webp)
Page 9 of 14
Part number 43028
|
Version 1.0
|
Effective 31JUL2020
© 2020 Codex DNA, Inc. All rights reserved.
Figure 3: Shipping screw location (shown with head cover removed)
13. After removing the power cord from packaging, connect
it to the power inlet on the system cart located on the
right rear side panel.
14. Check that the power switch located above the power
cord inlet is in the “off” position (0).
15. Connect the system to a grounded power outlet.
16. Switch on the power at the power inlet.
17. The system will boot up, perform a self-check, and
present the BioXp™ 3200 home screen once complete.
18. The system is now ready to operate.
Ethernet setup
The BioXp™ 3200 system requires an internet connection.
To set up the internet connection:
1. Connect an ethernet cable from the access panel located
on the right side of the unit to an available ethernet
connector located in your lab.
1. The BioXp™ system will need port 80/443 connections
opened on your network to access the Internet.
1. The following websites must be whitelisted for the BioXp™
system to communicate with the Codex DNA server and
for remote troubleshooting activities:
drive.google.com
,
logmein.com
and
codexdna.com
System startup self-check
The BioXp™ 3200 system contains a self-check process which
verifies the correct functionality of motion control and other
internal systems of the instrument. You may notice the system
performing these tests when it starts up. If the system is
unable to perform a successful self-check, it will alert the user
to the problem and will prevent the system from operating until
the problem has been addressed.
The self-check process has been carefully developed to
ensure that the BioXp™ system is operating within optimal
calibration ranges, and that all systems are functioning
correctly.
Please check that the globe in the upper right is spinning;
this indicates that the system is connected to the internet
and to the Codex DNA server. If the globe is grayed out,
please check that internet connection requirements are met.
If the issue persists, contact Codex DNA customer service at
.