CODEX DNA BioXp 3200 system Installation Manual Download Page 1

Page 1 of 14
Part number 43028  

 Version 1.0  

|

  Effective 31JUL2020

© 2020 Codex DNA, Inc. All rights reserved.

BioXp™ 3200 system 
Remote installation and sign-off

Summary of Contents for BioXp 3200 system

Page 1: ...Page 1 of 14 Part number 43028 Version 1 0 Effective 31JUL2020 2020 Codex DNA Inc All rights reserved BioXp 3200 system Remote installation and sign off ...

Page 2: ... installation point Internet connection It is essential to provide a fast internet connection for the BioXp system to support the electronic transfer of job information between Codex DNA and your lab Please note that the system can run locally for installations where there are internet restrictions for instruments such as the BioXp system However we strongly suggest you have your BioXp system conn...

Page 3: ...nded below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 3 5 P ________ V1 80 MHz to 800 MHz d 3 5 P ________ E1 800 MHz to 2 5 GHz d 3 5 P ________ E1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters ra...

Page 4: ...stem Allow 10 cm of clearance on both sides of the instrument for proper ventilation Make certain that the power cord connection point is free and clear of obstructions so that it is easily accessible The intended location must be stable so that the instrument will not move or vibrate excessively when in use Avoid dusty areas Avoid areas where the instrument would sit in direct sunlight or near a ...

Page 5: ...mage to the equipment and or personal injury 1 Unfasten the four draw latches on the top and bottom of the crate Lift and remove the lid and foam cap Lift the sides clear of the instrument 2 Remove the foam packaging materials from the top of the instrument 3 Carefully coordinate lifting the BioXp system off of the pallet with at least two people using proper lifting technique 4 Making certain the...

Page 6: ... from packaging connect it to the power inlet located on the right rear side panel of the instrument 6 Check that the power switch located above the power cord inlet is in the off position 0 7 Connect the instrument to a grounded power outlet Note We recommend connecting the instrument to a 1500 VA 900 W uninterruptible power supply ...

Page 7: ...rom any wall on a side to provide adequate ventilation The unit may be placed with the back against a wall The instrument must sit on a clean dry flat and level surface capable of holding the weight of the machine The intended location must be stable so that the instrument will not move or vibrate excessively when in use Avoid dusty areas Avoid areas where the instrument would sit in direct sunlig...

Page 8: ...e lift the unit clear of the base and set on the movable cart 6 Move the system using the cart to the prepared site 7 Codex DNA recommends keeping the shipping materials until installation and acceptance testing have been completed 8 Remove tape holding lid and open lid 9 Remove any tape internal to the system holding the thermal cycler door closed 10 Remove the two white covers for the chillers f...

Page 9: ...twork to access the Internet 1 The following websites must be whitelisted for the BioXp system to communicate with the Codex DNA server and for remote troubleshooting activities drive google com logmein com and codexdna com System startup self check The BioXp 3200 system contains a self check process which verifies the correct functionality of motion control and other internal systems of the instr...

Page 10: ...and plate location calibrations 3 The following operations are performed during the DEMO job Plate press functions Output plate Reagent plate Construct Plate Construct plate functions Placement of cover onto stand and back Movement of construct plate to magnet stand and back Thermal cycler Door open and close Cover movement Cover reagent chiller from storage location to reagent plate and back Cove...

Page 11: ...red at 4 C Load the recovery plate onto the recovery chiller with the notch in the upper left corner Load the Oligo Vault plate so that the notch is positioned in the upper left corner of the thermocycler 6 Briefly spin the DNA purification strip for 1 to 3 seconds Load the black DNA purification strip into the 1 position left most with the strip pinhole closest to the front of the instrument 7 Se...

Page 12: ... recovery plate shown in figure 6 The quality of the BioXp fragments is a key indicator of overall assembly process and successful results ensure the BioXp system is correctly installed and is working properly Plate 1 1 4 1418082298 Order 4 Date December 08 2020 Customer Codex DNA Project STOCK 32 Sequences Control 31 1 2 3 4 5 6 7 8 9 10 11 12 A B C D E F G H A1 1416 bp control B1 1416 bp control...

Page 13: ...rights reserved These gel images show an example of a completed 45PC run with the assembled fragment of interest clearly distinguishable and similar quantity of product Verify the intensity of the bands on the gel is similar to the above image All lanes verified Yes No Initials __________ ...

Page 14: ...is instrument and gives Codex DNA Inc the approval to invoice the customer via the agreed upon terms of sale or placement Instrument install date Serial number Customer name Company or institution Customer address Phone number Fax number Email address Application Customer signature Date Internet connection mode Local Online Email a copy of the completed remote installation and sign off along with ...