where :
-
Modulation Frequency =
30 to 45Hz
-
Mark/Space Ratio
=
0.9 to 1.1
-
Saturation Count
=
50 maximum for all analysers
Should the values be outside of the above ranges, the Data valid LED will
extinguish, and the fault condition be displayed – see Data Valid LED Out
for details.
If one or more fault conditions occur, the data valid LED on the front panel will
extinguish, the data valid relay will operate and the instrument will automatically
enter the diagnostic mode to display the fault condition - see section
Fault Condition.
6.3.2
Data Valid LED Out