Information screen
The list number must be different than 255255255 but in case it is the same will appears a program
error. To solve the problem it is necessary to reprogram the equipment. If the list file is not available,
contact Technical assistance.
If the number that appears is correct, find in the code list a reserved number (in case is not available,
contact Technical Assistance) and enter it in the codes screen.
Code screen
Code list
Once the reserved code is entered the light will turn green and the machine will emit 4 long beeps,
identifying that the code is correct.
IMPORTANT! Each code list is UNIQUE, belonging only to one equipment and can not be applied to
another.
3.2.9 - Equipment does not suck
If at the beginning of a treatment the equipment reaches the programmed vacuum even when the
hand piece is not placed on a body, it is because there is an obstruction in suction circuit.
The first step in order to solve the problema, is to check that no hose is bent in the rear part.
Summary of Contents for Cooltech
Page 1: ...Cooltech SERVICE MANUAL Service Manual Cooltech V1 1 ...
Page 5: ... A4 01454 x2 A4 01453 A4 01454 A4 01455 A4 01455 A4 01455 x2 x2 x2 ...
Page 7: ...Packing layout A4 02266 A4 01389 x2 A4 01591 A4 01391 A4 01415 ...
Page 8: ......
Page 12: ...1 1 2 Vacuum system ...
Page 37: ...Power supply mounting screws For better access to the screws ...
Page 50: ...Power supply screws For better access to the screws ...