6
Quick Bump Test
The quick bump test will only work if the alarm is set to the factory default of 5,000ppm.
1.
Turn the device on and wait for it to complete the 5 second warm up
cycle.
2.
Place the top end of the monitor close to your mouth and
breathe into the holes repeatedly.
3.
The CO2 in your breath will cause the CO2 in the case to
increase. Continue to blow until the alarm sounds. You can verify
this on the screen.
4.
Stop blowing. As your breath inside the case is displaced, the
CO2 level will return to normal.
5.
If the monitor does not pass the bump test, do not use it
. First,
perform a full bump test. If the full bump test fails, contact
CO2Meter support at (386) 256-4910 or e-mail us at
MENU OPTIONS
Press the Menu button
to enter the menu mode. There are 5 menu pages: Alarm,
Temperature units, Calibration, Altitude and Data Logging. Press the Set/Home button
to
return to normal display.
Note: If the monitor is on and the screen is black, press any button to access the menu options.
1. Settings
The device is equipped with a 30 sample rate default setting, which has the function to set record
periods as shown below. Pre-set and default record periods are displayed, however to change
default alarms see as follows.
1.
Press Menu
once to select the record period.
2.
Press Next
to change the default setting.
3.
Press Set
to save and return to normal display.