10. Customizing the Settings 

10.1 Select Temperature Units:

 Press the Up button to switch between ºF & ºC: 

 

10.2 Select Barometric Units:

 Press the Down button to switch from hPa to inHg 

 

10.3 Diagnostic Test:

 Test the communication link between the SEU & RDU.

 

 

1. Press Mode until the “DIAG” icon flashes 
2. Press Enter. The five LED’s will blink and the buzzer will sound on the SEU and 

RDU simultaneously. 

10.4 Set AL1 Alarm Level: 

(only if Alarm level is unlocked - see 9.1)

 

1. Press Mode until the “O2” & “AL1” icons flash 
2. Press Enter, the “AL1” icon shows on LCD. Default AL1 = 23.0% 
3. Press Up/Down to go through AL1 levels: 21%, 21.5%, 22%, 22.5%, 23%, 
23.5%, 24%, 24.5% and 25% 
4. Press Enter to save or press Mode to exit without saving. 

10.5 Set AL2 Alarm Level: 

(only if Alarm level is unlocked - see 9.1)

 

 

1. Press Mode until the “O2” “AL2” icon flashes 
2. Press Enter, the “AL2” icon shows on LCD. Default AL2 = 24.0% 
3. Press Up/Down to go through AL2 levels: 21%, 21.5%, 22%, 22.5%, 23%, 

23.5%, 24%, 24.5% and 25% 

4. Press Enter to save or press Mode to exit without saving. 

10.6 Set AL3 Alarm Level: 

(only if alarm level is unlocked - see 9.1)

 

 

1. Press Mode until the “O2” “AL3” icon flashes 
2. Press Enter, the “AL3” icon shows on LCD. Default AL3 = 25.0% 
3. Press Up/Down to go through AL3 levels: 21%, 21.5%, 22%, 22.5%, 23%, 

23.5%, 24%, 24.5% and 25% 

4. Press Enter to save or press Mode to exit without saving. 

Note: The 3 alarm levels cannot overlap. 

10.7 Calibration:   

Zero and Span Calibration should both be performed at least annually. They can 
be performed onsite or the SEU can be returned for factory calibration. Check your 
state or local code for calibration schedule requirements in your jurisdiction. 
What you will need: a test gas cylinder of pure nitrogen (0% O2), a cylinder of 21% 
oxygen, a regulator, tubing, and a small piece of masking tape to cover the buzzer 
hole on the bottom right corner of the SEU. If you are calibrating in situ, first inform 
occupants to ignore the alarms while calibration takes place. 
 

Part 1: Zero Calibration: 

 

1. Hold the tubing from the Nitrogen (0% O2) gas cylinder to the oxygen sensor 

opening on the top right-hand corner of the SEU. 

2. Open the gas regulator. Set the flow rate between 0.15-0.50 liters per minute.   
3. Wait until the O2 reading on the SEU settles consistently at or near 0% O2 on 

the screen. All alarms will be triggered. 

4. Press the Mode button 5 times. The “O2” and “CALI” icons will flash. 
5. Press Enter to view the calibration settings. The words “O2” and “ZEro” will 

begin flashing on the LCD. 

6. Press Enter again to begin calibrating. “CALIBRATING” will begin flashing. 
7. After approximately 3 minutes, the LCD will display either “PASS” or “FAIL” 
8. If the LCD reads “FAIL”, repeat the steps again. If it displays “PASS”, press 

Enter. The display should now show 0% 

9. Remove the nitrogen gas. 
10. Press the Reset button at the bottom right-hand side of the SEU front cover. 

 

Part 2: Span Calibration: 

1. Hold the tubing from the 21% oxygen gas cylinder to the oxygen sensor 

opening on the top right-hand corner of the SEU. 

2. Open the gas regulator. Set the flow rate between 0.15-0.50 liters per minute.   
3. Wait until the O2 reading on the SEU settles consistently at or near 21% O2 on 

the screen. 

4. Press the Mode button 5 times. The “O2” and “CALI” icons will flash. 
5. Press Enter to view the calibration settings. The words “O2” and “ZEro” will 

begin flashing on the LCD. 

6. Press the Up or Down arrow key to change from “Zero” to “SPAn”. 
7. Press Enter again to begin calibrating. “CALIBRATING” will begin flashing. 
8. After approximately 3 minutes, the LCD will display either “PASS” or “FAIL” 
9. If the LCD reads “FAIL”, repeat the steps again. If it displays “PASS”, press 

Enter. The display should now show 21% O2. 

10. Remove the oxygen gas. 
11. Press the Reset button at the bottom right-hand side of the SEU front cover. 
12. Remove tape over the alarm buzzer holes. 

 

10.8 Recover Factory Settings: 

If calibration or changing custom parameters has caused the RAD-0012 to behave 
incorrectly, use the ReFactSet (Recover Factory Settings) function to return the 
unit to its original condition at shipping.   

 

1. Press Mode until the “ReFactSet” icon flashes. 
2. Press Up/Down button to choose either “Yes” or “No”. 
3. Press Enter to save or press Mode to exit without saving. 

 

11. Product Care 

To insure you receive the maximum benefit from using this product, please 
observe the follow guidelines. 
1.

 

Repair - Do not attempt to repair the product or modify the circuitry by yourself. 
Please contact CO2Meter directly if the product needs servicing, including the 
replacement or calibration of sensor See section 15 for technical support 
contact information. 

2.

 

Cleaning - Disconnect the power before cleaning. Use a damp cloth. Do not use 
liquid cleaning agents such as benzene, thinner or aerosols, as these will 
damage the device. 

3.

 

Maintenance – We recommend testing the communication between the SEU 
and RDU under ‘DIAG” function. If these five LEDs blink and the buzzer of SEU 
and RDU sound simultaneously, it indicates that SEU and RDU work normally. 

12. Specifications 

 Oxygen & Temperature Sensor Specifications: 

Oxygen Sensor Specifications

 

Measurement Range  0 - 25% display   
Display Resolution 

0.1%   

Accuracy 

Better than 2%FS 

Pressure 
Dependence 

Auto pressure compensation, built in barometer. (500
1200 mbar) 

Response Time 

O2: <2min by 90% 

O2 AL1   

21%, 21.5%, 22%, 22.5%, 23%, 23.5%, 24%, 24.5% and 
25%. 
The default AL1 is 23.0%. 

O2 AL2   

21%, 21.5%, 22%, 22.5%, 23%, 23.5%, 24%, 24.5% and 
25%. 
The default AL2 is 24.0% 

O2 AL3 

21%, 21.5%, 22%, 22.5%, 23%, 23.5%, 24%, 24.5% and 
25%. 
The default AL3 is 25.0%. 

Sound Alarm 

80db@10cm 

Warm-Up Time 

< 60 seconds at 72°F (22°C) 

 

Temperature Sensor Specifications: 

Temperature Range 

32°F to 122°F (0°C to 50°C) 

Display Resolution 

0.1°F (0.1°C) 

Display Options 

°F /°C 

Accuracy  

±2.7°F (±1.5°C) when O2 concentration is above first 
alarm level 

Response Time 

20-30 minutes (case must equalize with environment) 

Operating Conditions: 

Operating 
Temperature 

32°F to 122°F (0°C to 50°C) 

Humidity Range 

0 ~ 95% RH non-condensing 

Storage Conditions:

 

Storage Temperature

 

-4°F to 140°F (-20°C to 60 °C)

 

 Power Supply   

Power 
Supply 

DC   

9~32VDC (12~32VDC recommended), 2A. 

AC adapter 

Input: 100~240 VAC,50/60Hz, 0.6A 
Output: 12VDC, 2000mA. 

Battery   Voltage 

6VDC (5.4V~7.0V), 
recommended capacity is 12AH 

 Relay Outputs

 

Relays 1-3 

Peak current for each relay < 2A @ 30VDC or 250VAC 
Relay configuration: SPDT 

 Weight and Dimensions

 

SEU 

Weight 1 

lb. 

Dimensions (LxWxD) 

6.69 x 4.96 x 2.73 inches 

RDU 

Weight 0.4 

lb. 

Dimensions (LxWxD) 

5.51 x 3.54 x 1.89 inches 

 

13. Safety Notes 

Warning: Your safety is very important to us. To ensure to use the product correctly 
and safety, please read these warnings and the entire User Manual before using the 
product. Otherwise, the protection provided by the equipment may be impaired. 
These warnings provide important safety information and should be observed at all 
times. 
1. Please handle the device carefully; do not subject the product to impact or 

shock. Otherwise, this may cause the accuracy drift. 

2. Do not place the unit near a heat source. Heat can cause distortion of the unit, 

which may result in an explosion or fire. 

3. Do not touch the exposed electronic circuitry of the device under any 

circumstances, as there is the dangerous of electric shocks. 

4. Please take care of cable connection between SEU and RDU. Make sure the 

cable from SEU is connected into the INPUT port of RDU. 

5. Please ensure the external power supply is normally supplied to ventilation fan 

while the relay is working. If there has no normally power supply to the fan, the 
relay will not work, which may result in potential danger with lower O2 
concentration in confined space. 

14. Fault Codes &Troubleshooting Guide 

This section includes a list of Frequently Asked Questions for problems you may 

encounter with the RAD-0012 O2 Monitor.   

No 

LCD 

Fault 

Icon 

Description  

(of the fault) 

SEU 

Indication 

RDU 

Indication 

Suggested Actions 

1 Er3 

 
The ambient 
temperature has 
exceeded the 
temperature range 
32°F to 122°F 
(0°C to 50°C) 
 
 

“Er3” flash, 

Fault LED 

blink    

“Er3” flash, 
Fault LED 
blink 

This error will 
disappear when the 
temperature returns to 
the range between 
32°F to 122°F (0°C 
and 50°C) 

2 Er5 

EEPROM System 
Problem 

“Er5” flash, 

Fault LED 

blink 

“Er5” flash, 
Fault LED 
blink 

Power on again or 
press reset button, if 
the “Er5” always 
appear, please contact 
with the local dealer. 

3 Er7 

Internal Data 
Transmission Error 

“Er7” flash, 

Fault LED 

blink, 

“Er7” flash, 
Fault LED 
Blink 

Check the RJ45 plug is 
connected into the 
INPUT port of RDU, if 
the “Er7” displays on 
the RDU only. 
Press reset button on 
SEU or power on 
again 

15.

 

Support & Warranty 

Contact us: We’re here to help!

 

If the troubleshooting guide above doesn’t help you solving your problem or for more 