Differential Pressure Range 

0-50”  H

2

O to 0 -100 PSID 

 

 

Leakage 

None Hi to Lo 

 

 

Pressure (Ratings) 

 

 

 

 Max 

Working 

See Specifications   

 

 

 

 

 

 

Gauge Accuracy 

2%   

ASME B40.100 GRADE B 

 

Operating Temperature (Max.) 

-20

 F   -  150

 F 

 

 

ELECTRICAL: 

 

 

 Min 

Typ 

Max 

 

 

Transmitter Accuracy (FSR) 

 

 

2% 

Upper 80% of Full Scale 
Range 

 

Supply Voltage (3) (Vdc) 

8  

 

28  

Pin 3  Reverse Polarity 
Protected 

 

 

Output Current (ma) 

 

 

 

  

Zero Floating (2) 

4.0 – 20.1 ma 

4.0 – 21.0 

4.0 – 22.0 

Pin 2 

 

 

 

Zeroed (1 connected to 2) 

 8   

 

 

 

Voltage (Pin 2 to 1) 

4.8   6.3 

 

 

 

Zero Time (seconds) 

2    

 

 

 

Max Loop Resistance (ohms) 

  

1000 

 

 

 

Max Loop Resistance Formula 

((Vs – 8)*1000)/ 20  

 

 

INTERFACE:

 

 

Electrical: 

 

 

 

 

Connections: 

4 Position Terminal Strip; ½” NPT Conduit 
1= Rtn,    2= Zero,     3 =  8-28 Vdc In   4= Chassis 

22 Awg – 16Awg Wire 

 

 Environmental 

Rating: 

NEMA 4X 

 

 

  

Certifications: 

 CSA  (Canadian & US Standards; Division 2 Locations) 

 

Electrical Configuration: T or W 

Model 140/141 End Connected 

With Transmitter 2 ½” Dial 

Electrical Configuration: T or W 

Model 140/141 End Connected 

With Transmitter 4 ½” Dial 

W option=   Class I, Div. 2, Groups A, B, C,D 

               Class II, Div 2, Groups F & G  

CMC TECHNOLOGIES 

P

TY

 L

IMITED

 ACN: 47 085 991 224    ABN: 47 085 991 224 

 

Phone: 

 

+61 2 9669 4000   

 

 

 

Unit 19, 77 Bourke Rd 

Engineering &

 Fax: 

 

 +61 

9669 

4111 

    Alexandria 

Industrial

  

Email: 

 

[email protected]

  

 

NSW  2015 

Instrumentation

 Website:

 

www.cmctechnologies.com.au

  

 

AUSTRALIA 

 

Reviews: