background image

4   |   

BT21i   Current   sensor   User’s   Guide      

Technical   Specifications   

   

Sensor   kind   

Analog,   generates   an   output   voltage   between      
-­‐7   and   7   

   

Measurement   range      

-­‐   5   A   ..   5   A      

Resolution      

3.8   mA   

Sensitivity   

1.28   V/A   

Calibration   function   

I   (A)   =   0.78125   *   V

out

   (V)   –   0.0047      

Shunt   

   

0.04   

Ω

      (2   Watt)   

Input   impedance   to   ground

   

each   terminal   400   k

Ω

      

Input   offset   current   error

   

typical      ±   8   mA      

Common   mode   input   voltage   error

    typical   1.5   mA/V      (0   –   500   Hz)      

Non-­‐linearity

   

<      0.001   %   

Slew   rate   

   

3   V/

µ

s   (maximum   output   voltage   variation)   

Bandwidth   (small   signal)

   

120   kHz   (-­‐3dB)   

Maximum    common    input    voltage      
Maximum   current

   

±   50   V   (max.   voltage   related   to   ground)   
7   A   

Supply   voltage

   

5 V DC 

Supply   current   

typical   23   mA   

Connection   

IEEE1394    connector    for    BT-­‐IEEE1394    sensor   
cable.    Sensor    cable    not    delivered    with    the   
sensor.   

 

 

Warranty:      

The    Current    sensor    BT21i    is    warranted    to    be    free    from    defects    in    materials    and   
workmanship   for   a   period   of   12   months   from   the   date   of   purchase   provided   that   it   
has   been   used   under   normal   laboratory   conditions.   This   warranty   does   not   apply   if   
the   sensor   has   been   damaged   by   accident   or   misuse.      

   

Note:

   

This   product   is   to   be   used   for   educational   purposes   only.   It   is   not   appropriate   

for   industrial,   medical,   research,   or   commercial   applications.             

   

   

 

 