BT24i   CO

2

   Gas   Sensor   User’s   Guide      

Technical   Specifications   
   

Sensor   kind   

Analog,   generates   an   output   voltage   between   0   -­‐   5   V      

Measurement   range

   

0   ..   5000   ppm   (0   ..   0.5   %)   at   1   atm   

Resolution   using   12   bit   AD   converter   

2.44   ppm   CO

2

   

Typical   resolution   

20   ppm   

Calibration   function   

CO

2

   (ppm)   =   2000   *   V

out   

(V)   

Live   calibration   by   calibration   button   

in   fresh   air   ~400   ppm   

Accuracy   (at   standard   1   atm)   

100   ppm   in   the   range   of   0-­‐1000   ppm   

10   %   of   reading   in   the   range   of   1000-­‐5000   ppm   

Warm   up   time   

90   seconds   (maximum)   

Response   time   

90%   of   full-­‐scale   reading   in   60   s   

Normal   operating   temperature   range   

20   ..   30   °C      

Operating   humidity   range   

5   ..   95%   RH   (non-­‐condensing)   

Connection

   

IEEE1394   connector   for   BT-­‐IEEE1394   sensor   cable.      
Sensor   cable   not   delivered   with   the   sensor.      

   

   

   

   

   

   

   

   

   

   

   

   

Warranty:      

The   CO

2

   Gas   sensor   BT24i   is   warranted   to   be   free   from   defects   in   materials   and   workmanship   for   a   

period    of    24    months    from    the    date    of    purchase    provided    that    it    has    been    used    under    normal   

laboratory   conditions.   This   warranty   does   not   apply   if   the   sensor   has   been   damaged   by   accident   or   

misuse.      

   

Note:

   

This    product    is    to    be    used    for    educational    purposes    only.    It    is    not    appropriate    for    industrial,   

medical,   research,   or   commercial   applications.      

   

       

   

 

 