We firstly need to capture images before adjus�ng grayscale. To adjust
grayscale range is for the background and the brightness of bands to be
best to watch. The grayscale range is from 0 to 65535.
The lower value means that the background is darker (sugges�on: not to
adjust). The lower is the high value, the brighter band is. You can control
the brightness of band by decreasing the value.
We only turn the High value down.
D. Preview and Capture
Display se�ng window
Adjust the grayscale range
Auto save file path
Click this icon to select folder path which is automatically
saved, shown as below:
It will be automa�cally saved into this folder when picture is taken in
every �me, and the customized file name , the date and �me of the
shoo�ng will be displayed , shown as below:
Turn on/off the preview mode: The image will be present-
ed based on parameters such as aperture, light source and
exposure �me.
I
mage capture manually: parameters are same as current
se�ngs when shoo�ng
Auto Exposure: The system automa�cally calculates the
exposure �me, the aperture, focus length and other par-
ameters are consistent with the lightness of samples.