background image

1000 ROUND - 1100 RECTANGULAR BOILING BATHS
CLIFTON RANGE® - 3 YEAR WARRANTY 

2

Dear Customer

The Clifton Range® is part of Nickel-Electro Ltd a family firm based in Weston-Super-

Mare which was incorporated as a limited company in 1941 but its roots can be traced 

back to 1935 when the business first started. Now in its 3rd generation of family 

members, the company prides itself on being a strongly established, independant 

British manufacturer. 

Thank you for purchasing this piece of Clifton Range® temperature control equipment. 

To get the best performance from your equipment and for your own safety please read 

these instructions carefully before use. 

GENERAL NOTES

1.  This product is designed for laboratory use only. Always follow good laboratory 

practice.

2.  Fill the tank prior to connection to electrical supply.

3.  Use caution when topping up/draining the tank. If this product is not used in 

accordance with these instructions then basic safety protection may be affected.

4.  The mains supply cord fitted to this product is heat resistant and should be 

replaced with an equivalent type by a qualified electrician.

5.  Ensure that the power supply has a safety earth (ground) terminal.

6.  Connect to a mains power supply with RCD protection if possible.

7.  Ensure that the mains switch and power supply connector are accessible during 

use.

8.  Before using any cleaning or decontamination method please refer to the 

Maintenance and Cleaning section to ensure the proposed method will not 

damage the unit. 

9.  Connect only to a power supply with the corresponding voltage to that specified 

on the rating label positioned on the rear of the unit. Ensure substances being 

heated present no risk of a hazard (explosion, implosion or release of toxic or 

flammable gases) or that these have been addressed.  When heating substances 

where liberation of gases occurs suitable extraction should be used.

10.  Use only liquids specified within this Instruction Manual within their specified 

temperature range.

11.  Drain before moving the bath.

12.  Over voltage category II (IEC 644) and Pollution Degree 2 (IEC 644).

13.  The 1000 and 1100 series are fitted wth a non-resetting overtemperature device 

which will operate once. The heater is capable of withstanding boil dry situations.

Summary of Contents for 1000 Series

Page 1: ...High quality products at affordable prices 1100 SERIES RECTANGULAR BOILING BATHS 1101E 1104E 1102E 1105E 1000 SERIES CIRCULAR BOILING BATHS 1026E ...

Page 2: ......

Page 3: ...s heat resistant and should be replaced with an equivalent type by a qualified electrician 5 Ensure that the power supply has a safety earth ground terminal 6 Connect to a mains power supply with RCD protection if possible 7 Ensure that the mains switch and power supply connector are accessible during use 8 Before using any cleaning or decontamination method please refer to the Maintenance and Cle...

Page 4: ...packaging and retain over the warranty period Contents consist of Bath Lid with concentric ring sets Power lead Instruction manual Remove any protective vinyl coating prior to use Place the lid onto the bath Fit the concentric ring sets Fit the power lead into the socket at the rear SAFETY Do not touch any electrical contacts or open any closure panels RISK OF ELECTRIC SHOCK POWER LEAD AND CONNECT...

Page 5: ...ATING INSTRUCTIONS BOILING 1 Always ensure the boiling bath is filled with water and the correct water level is maintained Ensure there is a slow running supply of water topping up for losses due to steam 2 Switch the mains electrical supply on and allow the water to boil 3 For temperatures above 60 C the bath should be used with a lid to achieve and maintain boiling DO NOT SWITCH ON OR USE UNLESS...

Page 6: ...high concentrations of halogens particularly chloride Halogen deposits may show as rust which can be cleaned off with nitric acid 10 on a cloth WEAR PROTECTIVE EQUIPMENT The Clifton Range bath should be emptied at the end of each day then for interior surfaces WASH RINSE DRY Constant Level Device Please ensure that the above cleaning process includes the outlet paying particular attention to flush...

Page 7: ...im must be in writing and give the serial number and description of the goods order number and date of delivery and will not apply where any names or serial numbers or other information which may be attached to or inscribed upon the goods have been removed covered up or defaced in any way 6 Any goods or parts thereof which may require repair or replacement shall be repaired or replaced at the disc...

Page 8: ...0 and 1100 Series 1000W 240V Element EE0987 1 1000 and 1100 Series Plug and Lead Set EX0621 1 1000 and 1100 Series PORTABLE APPLIANCE TESTING These tests should be conducted by a qualified person DO NOT PAT test the boiling bath unless it contains water DO NOT Flash Test ACCESSORIES FOR THE 1000 AND 1100 SERIES OF BOILING BATHS ER 1 Energy Regulator Adds level of boiling control to circular and re...

Page 9: ...e following standards BS EN 61010 1 BS EN 61010 2 010 Safety requirements for electrical equipment for measurement control and laboratory use BS EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements Designed and Manufactured in the UK by Nickel Electro Limited Oldmixon Crescent Weston super Mare North Somerset BS24 9BL t 01934 626691 f 01934 630300 w www clifton...

Page 10: ...FINAL INSPECTION AND ELECTRICAL SAFETY TEST REPORT BX0853 Issue 13 April 2014 ...