1000 ROUND - 1100 RECTANGULAR BOILING BATHS
CLIFTON RANGE® - 3 YEAR WARRANTY
2
Dear Customer
The Clifton Range® is part of Nickel-Electro Ltd a family firm based in Weston-Super-
Mare which was incorporated as a limited company in 1941 but its roots can be traced
back to 1935 when the business first started. Now in its 3rd generation of family
members, the company prides itself on being a strongly established, independant
British manufacturer.
Thank you for purchasing this piece of Clifton Range® temperature control equipment.
To get the best performance from your equipment and for your own safety please read
these instructions carefully before use.
GENERAL NOTES
1. This product is designed for laboratory use only. Always follow good laboratory
practice.
2. Fill the tank prior to connection to electrical supply.
3. Use caution when topping up/draining the tank. If this product is not used in
accordance with these instructions then basic safety protection may be affected.
4. The mains supply cord fitted to this product is heat resistant and should be
replaced with an equivalent type by a qualified electrician.
5. Ensure that the power supply has a safety earth (ground) terminal.
6. Connect to a mains power supply with RCD protection if possible.
7. Ensure that the mains switch and power supply connector are accessible during
use.
8. Before using any cleaning or decontamination method please refer to the
Maintenance and Cleaning section to ensure the proposed method will not
damage the unit.
9. Connect only to a power supply with the corresponding voltage to that specified
on the rating label positioned on the rear of the unit. Ensure substances being
heated present no risk of a hazard (explosion, implosion or release of toxic or
flammable gases) or that these have been addressed. When heating substances
where liberation of gases occurs suitable extraction should be used.
10. Use only liquids specified within this Instruction Manual within their specified
temperature range.
11. Drain before moving the bath.
12. Over voltage category II (IEC 644) and Pollution Degree 2 (IEC 644).
13. The 1000 and 1100 series are fitted wth a non-resetting overtemperature device
which will operate once. The heater is capable of withstanding boil dry situations.