16
After the cuff pressure is
reached, the cuff begins to
deflate. As the cuff deflates,
deflation symbol “
“ appears
on the display. You will also
see the number decreasing
and Pulse Symbol “
“ flash
-
es on the display when pulse
is detected. (Fig 15)
Step5
- After completely deflation,
the monitor displays the
systolic pressure, diastolic
pressure, and heart rate.
(Fig. 16) Press the “ ” but-
ton to switch off or leave it
for 3 minutes for auto-shut
down.
Note:
It is extremely important that the cuff is at the same level
of the heart. If the wrist is relatively lower (higher) than the
heart, the blood pressure measured will be higher (lower)
than the actual value.
If a higher pressure value is needed, the monitor will stop
deflation and inflate again.
If you press the “ ” button during the measurement, the
monitor will turn off.
(Fig 16)